Ecg Installation Procedure For Systems With No Printer Installed - GE Venue R1 Service Manual

Table of Contents

Advertisement

P R E L I M I N A R Y
D
5771498-100, R
IRECTION
8-8-3

ECG Installation Procedure for Systems with No Printer Installed

8-8-3-1
Tools
Appropriate Phillips screwdriver
FRU Part #
Table 9-20
Refer to
8-8-3-2
Time Required
15 min
8-8-3-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-8-3-4
ECG Installation
1.) Remove the following covers: Lower Front eTower, Upper Front eTower, and MPB Front Metal
Door, Printer Insert Cover.
2.) Replace the printer insert cover with the one provided in the ECG installation kit.
6
EVISION
.
on page 9-19
Lower Front eTower Cover Removal Procedure
Upper eTower Front Cover Removal Procedure
MPB Front Metal Door Removal Procedure
Printer Insert Cover Removal Procedure
Figure 8-185 Replacing Printer Insert Cover
Chapter 8 - Replacement Procedures
V
™ S
ENUE
Power Shut Down
.
M
ERVICE
ANUAL
8-173