Physical Inspection - GE Venue R1 Service Manual

Table of Contents

Advertisement

P R E L I M I N A R Y
D
5771498-100, R
IRECTION
10-5-3

Physical Inspection

Table 10-8
Step
Item
1
Labeling
2
Scratches & Dents
3
Covers
4
Input Power
5
External I/O
6
Wheels and Brakes
11
Cables and Connectors
12
Shielding and Covers
13
Probe Holders
Power and System
14
Status Indicators
15
Battery
6
EVISION
Physical Checks
Verify that all Ultrasound system labeling is present and in readable condition.
Inspect the exterior for dents, scratches or cracks.
Where applicable, verify all covers are secured in place and are properly aligned with other covers.
Replace any covers that are damaged.
Refer to:
Mains Cable Inspection
Check all connectors for damage.
• Where applicable, check all wheels and casters for wear and verify operation of foot brake, to stop
the Ultrasound system from moving, and release mechanism.
• Where applicable, check all wheel locks and wheel swiveling for proper operation.
Check all internal cable harnesses and connectors for wear and secure connector seating. Pay
special attention to probe strain or bend reliefs.
Check to ensure that all EMI shielding, internal covers, air flow panels and screws are in place.
Missing covers and hardware could cause EMI/RFI problems while scanning.
Where applicable, inspect the Probe Holders for cracks or damage.
Check for proper operation of all Power and System Status Indicators.
Where applicable, check that the battery is not damaged, does not leak, does not emit an odor, and
is not deformed or discolored. Observe all warnings and cautions for battery handling, recharging,
storing, and/or disposal.
Chapter 10 - Care and Maintenance
Description
on page 10 - 8.
V
™ S
M
ENUE
ERVICE
ANUAL
10-9