Front End Connections; Application Error: "Abnormal - GE Venue R1 Service Manual

Table of Contents

Advertisement

P R E L I M I N A R Y
D
5771498-100, R
IRECTION
7-8-5
Application error: "Abnormal...behaviour"-->Front End Connections
Symptom: Auxiliary Display works (LEDs on). The application starts with an Error: "Abnormal system
behavior occurred"
Suggested solution: Reset front end connections
1.) Inspect the BIB (blue) cable, if damaged - replace the BIB board.
2.) Reset the BIB (blue) cable on the T-CFE side (disconnect and re-connect).
3.) Inspect and re-connect MPB to T-FEPS power cable. Replace if damaged.
4.) Check power cable to Frontend; If damaged - replace power cable
5.) Reseat all 4 T-TRX boards.
6.) Reload software
7.) Replace T-CFE board.
7-20
6
EVISION
Figure 7-41 Reset BIB cable
Figure 7-42 Reset T-FEPS power cable