Maintenance

Flow Cell Maintenance

Follow these instructions if you need to clean the
flow cell, unclog the flow cell, replace the inlet or
outlet tubing, or replace the flow cell.
Be extremely careful when working with the flow cell
and its fittings. Flow cells are considered expendable
and are not covered by Gilson's warranty if damaged
or broken during any maintenance procedure.
Cleaning the Inside of the Flow Cell
(Analytical, Preparative, Microbore,
and SFC only)
There are two ways to clean an active flow cell that
has become coated with light absorbing material:
• solvent cleaning
• acid cleaning