Maintenance

Fuse Replacement

To change a fuse, follow these steps.
1
Disconnect the power cord from the power
outlet and from the rear panel receptacle.
2
Locate the fuse drawer on the back panel.
3
Insert a small screwdriver into the notch
located below the fuse drawer.
4
Twist the screwdriver to open and remove the
fuse drawer. The fuse drawer contains one fuse
for a 90–120 voltage selection. It contains two
fuses for a 220–240 voltage selection.
5
Remove the old fuse(s) and insert the new
fuse(s). The replacement fuses must be 2.5 amp
"T" type fuses. The size of the fuses is 5 x 20 mm.
6
Insert the fuse drawer into its receptacle in the
detector.
A blown fuse may indicate the existence of
another problem in the instrument. If the
replacement fuses blow, don't try others. Contact
your local representative or Gilson. Refer to the
Before Calling Us
information on page 5-10.