Gilson 155 User Manual page 145

Uv/vis detectors
Table of Contents

Advertisement

Appendix
C-2
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
At a constant SENSITIVITY setting, the peak height
varies as the sample concentration changes.
• At the same SENSITIVITY setting, a sample with
an absorbance of greater than 1.0 produces a
peak height that is greater than full scale.
• At the same SENSITIVITY setting, full-scale
recorder pen deflection occurs when the sample
flowing through the cell has an absorbance of
1.0 AU.
• At a SENSITIVITY setting of 1.0 AUFS, a
sample with an absorbance of less than 1.0 AU
produces a maximum peak height that is less
than full scale.
C

Advertisement