Operation
Parameter
Automatic scan initiation
Initiation threshold
3-36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Third Display for Scan Wavelength
Mode
Following is a description of the parameters on the
third display for scan wavelength mode.
Description
When automatic scan initiation is "On" and you
press the Scan soft key, the scan is not performed
until the absorbance reading reaches the specified
limitation threshold, described below.
The Change soft key sets automatic scan initiation
to "On" or "Off".
When automatic scan initiation is "On" and you
press the Scan soft key, the detector will not
perform the scan until the absorbance reading
reaches the initiation threshold.
Output Channels
The following table lists what data are transmitted
via each output channel.
Channel 1: detection at monitor wavelength
(sensitivity 1)
Channel 2: detection at monitor wavelength
(sensitivity 2)
Channel 3: scan output; scan from wavelength 1 to
wavelength 2 (sensitivity 3)
3