Gilson 155 User Manual page 102

Uv/vis detectors
Table of Contents

Advertisement

Maintenance
To check for leaks:
1
To determine whether the fittings are leak-free,
use a syringe to initiate flow through the cell.
2
Check carefully along the sides of the cell for
beads of liquid. A flashlight may be helpful.
3
If leakage is detected visually or if the area
around the fitting feels cool to the touch,
remove and clean the flow cell as well as its
chamber. Then re-insert the flow cell and re-
tighten the fittings.
To replace the flow cell assembly, replace the
detector block, and close the detector door:
1
Insert the flow cell assembly into the detector
and tighten the flow cell assembly screws.
2
Replace the detector block and tighten its two
screws.
3
Close the detector's front door.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
4
4-11

Advertisement

Table of Contents
loading

This manual is also suitable for:

156

Table of Contents