Gilson 155 User Manual page 156

Uv/vis detectors
Table of Contents

Advertisement

Appendix
R
I
Wn=xx.xx
B
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Reads reference signal current. Returns: "n..n" in µFS units.
Disconnects and writes to the display. Or, reconnects the
display after it was disconnected. When specifying
parameters for the command:
n
- Line number (0 to n–1, n being the line number).
For instance, if you're writing to line 1, indicate 0
for n.
xx..xx - Alphanumeric string of up to 40 characters. The
string can be made up of letters (A–Z), numbers
(0–9), or special characters, such as these:
(!@$%^&*()_+–={}[]:;"<>,.?'/ and space)
If you specify more than 40 characters, the additional
characters are not shown on the display.
When the command is sent, the designated line is
disconnected from the detector's software. Lines may be
individually reconnected to the detector's software; see
syntax table below.
Following are several syntax examples for the buffered W
command:
Syntax example
Description
W0=xx..x
Disconnects and writes the character
string (xx..x) to line 1 of the display.
Wn=xx..x
Disconnects and writes the character
string (xx..x) on line n. For n indicate a
number from 0 to 3, with 0 being line 1 of
the display.
W=
Disconnects the whole display.
W0=
Disconnects line 1.
Wn=
Disconnects line n. For n indicate a number
from 0 to 3, with 0 being line 1 of the
display.
W
Reconnects all lines.
W0
Reconnects line 1.
Wn
Reconnects line n. For n indicate a number
from 0 to 3, with 0 being line 1 of the
display.
D