Single Wavelength Mode

In single wavelength mode, you can monitor a
sample at any wavelength from 190 to 700 nm, at 1
or 2 sensitivities. Plus, you can direct one output
channel to trace the percent of UV or visible light
transmitted through the sample.

Display For Single Wavelength Mode

Following is a description of the parameters on the
display for single wavelength mode. The absorbance
at the currently selected wavelength appears in the
upper left of the display. You use the display to set
the new wavelength, sensitivities for output
channels 1 and 2, and peak width.
Description
In the first line of the display, specify the desired
monitor wavelength. The default is 254 nm, and
the valid range is 190 to 700 nm.
Note: If the lamp saver option is enabled and you
indicate a wavelength not in the range of the
currently lit lamp, the detector lights the necessary
lamp. (For the UV lamp, the process of lighting the
lamp takes about 30 seconds.) Allow the baseline
to stabilize before using the new wavelength.
See page
3-19
for information on the lamp saver
option. By default, the lamp saver option is enabled.