Third Display For Setup Mode; Event Channels; Autorange Channels; Autorange Multiplier - Gilson 155 User Manual

Uv/vis detectors
Table of Contents

Advertisement

Operation
Parameter
Event channel(s)
Autorange
channel(s)
Autorange
multiplier
3-18
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Third Display For Setup Mode

Following is a description of the parameters on the
third display for setup mode. When you're finished
checking or changing parameters, press the Next
soft key to reveal the next display for setup mode.
Description
Set this parameter to tell the detector which output
channel(s) to use for marking event spikes.
Enter 0 (for no event mark), 1, 2, and/or 3. For
example, entering 2 activates channel 2; entering
123 activates all 3 channels.
Set this parameter to designate the output
channel(s) to autorange. When the detector signal
exceeds 100% of full scale, the autorange feature
attenuates the signal of the specified channel(s) to
a percentage of actual value. The output on
unspecified channel(s) is not affected.
Enter 0 (for no autoranging), 1, 2, and/or 3. For
example, entering 2 activates channel 2; entering
123 activates all 3 channels.
Use this parameter to designate the percent
attenuation of the signal for the specified autorange
channel(s).
For the multiplier, indicate a value from 0.1 to 0.5
(which is 10% to 50%). The default is 0.2 (which is
20%).