Gilson 155 User Manual page 107

Uv/vis detectors
Table of Contents

Advertisement

Maintenance
4-16
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
SFC or preparative SFC flow cell
To replace the flow cell window and other
components for the SFC flow cell, order the SFC
rebuild kit (part number 10043888). Instructions for
rebuilding the flow cell are supplied with the kit.
To access and remove the flow cell assembly:
1
Open the detector's front door.
2
Using a 5/32" Allen wrench, remove the two
screws on the front of the detector block and
move the block to the side.
3
Using a 7/64" Allen wrench, loosen the flow
cell assembly screws (refer to diagram on the
next page for location of the screws). Pull the
assembly towards you to remove it.
4

Advertisement

Table of Contents
loading

This manual is also suitable for:

156

Table of Contents