Troubleshooting
• Check for contaminated or "bleeding" column.
If necessary, regenerate or replace column.
• Check flow cell for contamination or damage.
Refer to
Flow Cell Maintenance
• A lamp may be weak. Refer to
Maintenance
on page 4-22.
• A lamp connector may be loose. Refer to
Maintenance
on page 4-22.
If none of the above fix the problem, contact the
Gilson Customer Service Department (refer to page
5-10 for contact information) or your Gilson
representative.
Peaks Too Broad
• Check peak width setting - it may be set too
high. Measure the narrowest peak; then set
new value.
• Check for contaminated column. If necessary,
regenerate or replace column.
Decreased Response to Known
Sample
• A lamp may be weak. Refer to
Maintenance
on page 4-22.
• Check connections to recorder. Refer to the
recorder's troubleshooting instructions.
• Verify sample concentration and solvent.
• Check flow cell for contamination. Refer to
Flow Cell Maintenance
Abnormally High System Pressure
• Check for blockage in flow cell or tubing.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
on page 4-2.
Lamp
Lamp
Lamp
on page 4-2.