Operation
Next, the display shows the conditions for the
last-used detection mode (single, dual, or scan
wavelength). The bottom line lists soft key
commands.
For example, if single wavelength mode was the
last mode used, you might see the following:
The detector stores all of your entries in its
permanent memory, so the detector always
"wakes up" in the last selected detection mode.
The detector also recalls the parameter values
you entered the last time you used the detector.
Note: If the last peak width value was 0.0
seconds, the detector will not default to this
value, it will default to 4.0 seconds. When using
a detector peak width value of 0.0 seconds the
peak width value must be reset when the
detector lamp is powered on or when modes
are changed in the detector software. (A peak
width setting of 0.0 seconds is only available in
single wavelength mode.)
Note: When you first use your detector, default
conditions and values are displayed. These
conditions and values will be applied until you
change them. See
Appendix A
defaults and the valid range of parameter
entries.
Note: The detector requires a 1-hour warm-up
period to achieve a stable baseline for use at
high sensitivities.
Note: If you cannot easily read the text on the
front panel, refer to
to adjust the view angle for the display.