Reset lamp alarm
The detector will flash the error and lamp status
indicators when the lamp has been in use for a
number of hours that exceeds the setting for the
lamp alarm. Do not turn the lamps off. Reset the
lamp alarm.
To reset the lamp alarm for the 156 Detector, refer
to the user's guide for the Gilson control software.
The software should include a command that resets
the lamp alarm. Or, the software should allow you
to send a GSIOC command to reset the lamp alarm.
See
Appendix D
for a listing of GSIOC commands.
To reset the lamp alarm:
1
Press Mode.
2
Press Setup to access setup mode.
3
Press Next until you see a screen similar to:
4
Disable the lamp alarm feature by typing 0 and
pressing ENTER.
5
Press ESC to exit setup mode.
6
Turn the lamps and MAINS switch off and
replace the lamp as described in this section.