Essential Setup - Gilson 155 User Manual

Uv/vis detectors
Table of Contents

Advertisement

Installation
2-4
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Essential Setup

Fuse installation
If you received the detector without the fuse drawer
and fuses installed:
1
Locate the accessory package containing the
fuse drawer appropriate for your line voltage.
Discard the other fuse drawer.
2
Locate the accessory package containing the 2.5
amp fuses.
3
Install the fuse(s) into the fuse drawer. The fuse
drawer for 90–120V accepts one fuse. The fuse
drawer for 220–240V accepts two fuses.
4
Insert the fuse drawer into its receptacle in the
detector. See the
rear panel diagram
2-3.
Output connections
Note: If the detector is part of a Gilson HPLC system,
you do not need to make output connections. Instead
make the GSIOC connection between the detector
and system controller as described in
Connections, page 2-6.
The detector has three output channels. All three
output channels are active. For more detail on the
output channels see
Detector Setup Parameters
Section 3,
Operation.
You can direct output from the detector to a chart
recorder or stand-alone integrator.
To connect an output channel, use a 5-foot shielded
Pomona recorder cable from your accessory kit.
One end of the cable has a dual-pronged plug. The
other end has two tinned wires.
2
on page
Optional
in

Advertisement

Table of Contents
loading