Gilson 155 User Manual page 82

Uv/vis detectors
Table of Contents

Advertisement

Operation
Indicate first set of conditions using single
wavelength detector mode
As stated earlier, the detector automatically copies
all detector parameters currently in memory to a
newly created program file. Therefore, you can set
up the initial run conditions using the single
wavelength mode instead of the file mode. When
you create the file, the detector automatically
assigns time 0.00 to each parameter. This saves you
time so you don't have to set up a timed event for
each of the initial conditions using the file mode.
To set up initial conditions using single wavelength
detection mode, do the following:
1
If the first display for single wavelength detection
mode is not shown, press Mode at the first display
for the current detection mode. You'll see:
2
Press Single. You'll see:
3
To set the monitor wavelength, press 254 on the
keypad, and then press ENTER.
4
To set the sensitivity, press 0.01, and then press
ENTER.