Command
Type
%
I
^
I
$
I
?
I
A
I
B
B
Dn
B
Fn
B
FE
B
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Description
Identifies the detector. Returns a character string that
identifies the model number and software version for the
detector, for example: "155V1.00"
Determines if the first display for single, dual, or scan
wavelength mode is shown. If the first display is shown,
returns "T" for top. Otherwise, returns the screen number
assigned to the display in the detector software.
Shuts down the detector and then turns on and resets the
detector to the last-used detection mode.
Identifies the current detection mode. Returns a number
from 0 to 2 where:
0 - Single wavelength mode
1 - Dual wavelength mode
2 - Scan wavelength mode
Reads active signal current and then returns "n..n" in µFS
units.
Causes the detector to beep for 0.1 second.
Turns off or on the front panel display depending on the
number (0 to 2) that you assign to n:
0 - Turns off display.
1 - Turns on display.
2 - Restores automatic mode if you turn off or on the
display using "0" or "1." With automatic mode, the
display is automatically turned off if there is no text on
the display.
For example, to turn off the display, send the following
command: D0
Starts file number (0–9) that you assign to n.
For example, to start file 1, send the following command: F1
Ends the execution of the file.