Gilson 155 User Manual page 63

Uv/vis detectors
Table of Contents

Advertisement

Operation
3-30
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Set parameters
Follow these steps to set the detection parameters:
1
To set the first monitor wavelength (240 nm),
press 240 on the keypad, and then press ENTER.
2
To set the second monitor wavelength (287
nm), press 287 on the keypad, and then press
ENTER.
3
To set the sensitivity at the first wavelength,
press 0.2, and then press ENTER.
4
To set the sensitivity at the second wavelength,
press 0.2, and then press ENTER.
5
Press Next to see additional parameters for dual
wavelength mode.
You'll see:
6
Assume that, at half height, the narrowest peak
in the run is 12 seconds in duration. If
necessary, press 12, and then press ENTER.
7
Set the ratio multiplier to 0.2, and then press
ENTER. You know from previous runs with this
sample that at this value all ratios are displayed
on-scale.