Discrete Sample Scan

To scan a single sample dissolved in a buffer, follow
these instructions.
1
Locate one 1/4-28 coupler and one male Luer
end-fitting from the accessory package. Make
the following connections:
• Attach the coupler to the detector outlet
• Connect the Luer end-fitting to the coupler
• Attach a syringe to the Luer end-fitting
2
Disconnect the detector inlet tubing from the
column. Immerse the end of the inlet tubing
into a pure buffer solution.
3
Draw the buffer solution through the inlet
tubing into the syringe.
4
From the first display for scan mode, press the
Zero soft key to perform a zero scan and to
reset the baseline of the trace at the monitor
wavelength to zero. The mobile phase spectrum
is automatically subtracted from the compound
spectrum during the scan.
5
Immerse the inlet tubing into the sample of
interest (in the same buffer).
6
Draw the sample solution through the inlet
tubing into the syringe. Be sure that the cell is
overfilled on both ends to eliminate bubbles
from the flow cell.
7
Press Scan to begin the sample scan.