Maintenance
4-6
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Analytical or preparative flow cell
To access and remove the flow cell assembly:
1
Open the detector's front door.
2
Using a 5/32" Allen wrench, remove the two
screws on the front of the detector block and
move the block to the side.
3
Using a Phillips screwdriver, remove the flow
cell assembly screw (refer to diagram below for
location of the screw). Then pull the flow cell
assembly towards you to remove it.
4
Refer to the following procedures and diagrams
when replacing the analytical or preparative
flow cell and/or tubing.
Flow cell assembly, assembled
Flow cell assembly, disassembled
4