Unit ID and Baud Rate Selection

On the detector's rear panel, use the SW1 selector
to choose a different unit ID and the SW2 to choose
a different baud rate.

Unit ID

The unit ID identifies the detector to Gilson software
packages that can issue GSIOC commands to the
detector via the GSIOC or RS-232 cable connection.
At the factory, Gilson set the unit ID to 16. There is
no need to change this number unless it is the same
as that assigned to another Gilson instrument that
is also connected along the GSIOC.
To change the unit ID:
1
Gently insert a small flat blade screwdriver into
the SW1 selector on the rear panel and turn it.
2
Align the white dot with one of the indicated
numbers. The unit ID is 10 plus the selected
number.
Note: The unit ID set for the detector in the control
software overrides the setting on the SW1 selector.
To reset the SW1 selector to the default setting of
16, set the SW1 selector to 6. Turn the detector
MAINS power switch off. Then turn the detector
MAINS power switch back on while pressing the
LAMP ON/OFF button.