Error Messages - Gilson 155 User Manual

Uv/vis detectors
Table of Contents

Advertisement

Operation

Error Messages

The following table lists and describes the error
messages that could appear during wake-up. To
remove the error message from the display so you
can remedy the problem, press OK.
Error Message
Invalid Lamp ID
Lamp or Home Failure
Low UV Energy
Low Visible Energy
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Possible Cause/Solution
The UV lamp may not be properly installed. See
Maintenance
on page 4-22. Contact Gilson if the problem
persists.
The UV lamp did not light, or there is an internal
misalignment. Follow the instructions in
Maintenance
on page 4-22. Contact Gilson if the problem
persists.
The UV lamp needs to be replaced, or there is an internal
misalignment. Perform
described on page 4-22. Contact Gilson if the problem
persists.
The visible lamp is disconnected or needs to be replaced, or
there is an internal misalignment. Perform
Maintenance
as described on page 4-29. Contact Gilson if
the problem persists.
UV Lamp
UV Lamp Maintenance
Visible Lamp
3
UV Lamp
as
3-15

Advertisement

Table of Contents
loading

This manual is also suitable for:

156

Table of Contents