Gilson 155 User Manual page 153

Uv/vis detectors
Table of Contents

Advertisement

Appendix
L0X
B
L1X
B
L
I
M
I
Pnn
B
Pnn=mmmmm B
P
I
D-6
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Sets the UV lamp hours to 0 (zero).
Sets the visible lamp hours to 0 (zero).
If both lamps are lit, returns "uuuu vvvv" where:
uuuu - UV lamp hours
vvvv - Visible lamp hours
Returns "Min=xxx Max=yyy" where:
xxx - Wavelength at which the minimum absorbance was
obtained during the last scan.
yyy - Wavelength at which the maximum absorbance was
obtained during the last scan.
If no scan has been performed, returns "Min=XXX
Max=XXX".
Sets parameter pointer to the specified parameter, where nn
identifies the code assigned to the parameter; see table on the
following pages.
Assigns a value (mmmmm) to a parameter (nn). If the value is
omitted, the parameter is set to zero. The table on the
following pages identifies the code assigned to each
parameter within a mode. It also lists the range of values that
can be entered. See
Section 3
additional information on the parameters within the modes.
Reads the value of a parameter. Returns "nn=mmmmm",
where "nn" identifies the parameter (see table on the
following pages) and "mmmmm" identifies the value. Use
the Pnn command to indicate the parameter to be reviewed.
and
Section 5
if you need