Gilson 155 User Manual page 84

Uv/vis detectors
Table of Contents

Advertisement

Operation
Indicate user events
Once you've selected the file number, you'll see:
In addition to the pre-stored time 0.00 events, you'll
specify events that do the following:
• zero the baseline at 0.01 min
• indicate a wait (for sample injection) at 1.0 min
• change the detection wavelength to 320 nm at
2.5 min
• change the sensitivity to 0.02 AUFS at 2.5 min
• zero the baseline again at 2.5 min
Zero the baseline
To add a timed event that zeroes the baseline:
1
Press Add. You'll see:
2
Press Misc. You'll see:
3
Type 0.01 and press ENTER.