Remove visible lamp
1
Locate the visible lamp.
2
Follow the visible lamp's wires to locate the
connector that links the wires to the internal
circuit board. Disconnect the connector from
the internal circuit board.
Note: If the connector was not secure in its
socket, you may want to secure the connector
and start the detector to determine if the loose
connector caused the problem.
3
Grasp the visible lamp and turn it until the flat
edge faces upward. Usually, this is just a slight
rotation of the lamp in either direction.
4
Remove the lamp by pulling it away from the
lamp housing.