Maintenance
4
Refer to the following diagrams when replacing
the flow cell and/or tubing.
Flow cell assembly, assembled
To replace the outlet tubing:
When replacing the outlet tubing, ensure that its ID
is appropriate for the SFC flow cell. See
for a list of replacement parts.
1
Remove the outlet tubing and fitting from the
flow cell assembly.
2
If necessary, position the fitting and ferrule on
the flow cell end of the replacement outlet
tubing. The tubing should extend 3 mm beyond
the ferrule. See diagram to right.
3
Seat the tubing into the top of the flow cell
assembly. Then tighten the fitting. The two
parts of the ferrule seat when you tighten the
fitting.
4
Route the free-end of tubing through the cutout
located at the lower left of the detector's front
panel.
5
If you're not replacing the inlet tubing, skip the
following instructions and proceed with the
check for leaks section on page 4-20.