Electrical Connection Diagram And Electrical Interfaces (Inputs And Outputs) - Honeywell enCore Series Manual

Gas quality analyzer
Hide thumbs Also See for enCore Series:
Table of Contents

Advertisement

5
5.6.5
Electrical connection diagram
and electrical interfaces (inputs and outputs)
The cables are connected to the board in the connection box via the plugs on the
edge of the board. All electrical interfaces (galvanically isolated) are located here.
The plugs are secured with snap-in hooks. The connections and interfaces are sup-
plemented and explained in figures and sections.
Connection
Interface
TB1
Power supply (see also section ).
TB2
Future extension.
Ethernet
TB3
electrically isolated, range: 100m ≙328.084 ft
RS485
TB4
(electrical isolation), range: 500m ≙1640ft
Digital
TB5
supplied by the device, max. 9 V to a potential-free contact or NAMUR sensor. e.g.
process and calibration gas pressure monitoring
4
digital
TB6
(1 breaker closed when de-energized, suitable for general alarms, 3 makers) (circuit
operates as a floating contact), max. 120 mA at 28.8 V (DC).
4 analog outputs 0/4...20mA (electrically isolated)
TB7
common return cable; short-circuit-resistant; R max. = 390 Ω
PROHIBITION INTERDICTION!
remove of the plug connection in the center of the connection
l
board. (see next figure)
retirer la fiche de connexion au centre de la carte de connexion. (voir
figure suivante)
Check that the connectors are secure!
Tighten the screws on any unused terminals!
Connect the cables and screens in compliance with the explosion pro-
tection regulations. In practice, the screens are not attached in explosion
hazard areas, but are only connected in the safe area.
The possible electrical connections for the GasLab Q2 are shown in the following
diagram:
GasLab Q2
GasLab Q2 after delivery and at the place of use
connection
10/100
port
1
and
inputs
1
and
2
outputs
(electrically
User Manual (N )
Mbit/s;
(TCP/IP;
port
2
(Modbus
(2
cores
each,
electrically
isolated)
common
Modbus
TCP)
serial)
isolated)
return
cable
81 of 315

Advertisement

Table of Contents
loading

This manual is also suitable for:

Gaslab q2

Table of Contents