GE Healthcare MAC ™ 1200/MAC ™ 1200 ST Resting ECG Analysis System (2018357-003 & 2013503-003 PWAs) Service Manual 2012250-095 Revision D...
Page 2
(PN 2018357-003 or 2013503-003) will render the MAC 1200 unit inoperable. MAC and MUSE owned by GE Medical Systems Information Technologies, Inc., a General Electric company, going to market as GE Healthcare. All other marks are owned by their respective owners.
Page 3
Compliance Compliance The country of manufacture can be found on the equipment labeling. The product complies with the requirements of standard EN 60601-1-2 “Electromagnetic Compatibility - Medical Electrical Equipment”. The safety and effectiveness of this device has been verified against previously distributed devices.
Page 4
Compliance For your notes Compliance-2 MAC 1200 resting ECG analysis system Revision D 2012250-095...
See the operator manual for the instructions necessary to operate the equipment safely in accordance with its function and intended use. This manual addresses both MAC 1200/1200ST units (international units) and MAC 1200 units (USA units). Information that is specific to either unit is identified.
The Basics: About This Manual Illustrations and Names All illustrations in this manual are provided as examples only. They may not necessarily reflect your monitoring setup or data displayed on your monitor. In this manual, all names appearing in examples and illustrations are fictitious. The use of any real person’s name is purely coincidental.
Modem. Additionally the MAC 1200 ST can run stress test operating mode (Option Code). To control stress test devices the MAC 1200 ST has a modified keyboard with additional keys for stress test. Revision D MAC 1200 resting ECG analysis system 1-5...
The following table gives an overview of the different device models of the MAC 1200 resting ECG analysis system. Device Device Manufacturer ID Keypad Labeling Mains Voltage Model MAC1200 GE Medical Icons International 230V MAC1200 ST Systems MAC1200 Europe 2 GE Medical Text...
The Basics: Mechanical Components Mechanical Components The main mechanical components comprise the upper and lower shell of the MAC 1200 resting ECG analysis system. The lower shell serves as a basic unit to receive the following assemblies. Line power input module with fuse ...
Functional Description: Basic Information Basic Information The block circuit diagram of the entire instrument in Chapter 1, “Block Circuit Diagram of Entire Instrument” and the functional blocks of circuit diagram sheets (P-plans) describe the individual functional blocks. Revision D MAC 1200 resting ECG analysis system 2-3...
Functional Description: Switching Power Supply Switching Power Supply The switching power supply generates the fundamental device voltage for the PCB Control CS_CI and for the PCB NC Battery Charge. It has the following input specifications. Input voltage range: 85 ... 264 Vac Line frequency range: 47 ...
Functional Description: PCB NC Battery Charge PCB NC Battery Charge The MAC 1200 resting ECG analysis system can operate on battery power. With a new, fully charged NC battery, up to 50 ECGs can be recorded in Automatic (12- lead) mode. The PCB NC battery charge controls and monitors the charging process, depending on the status of the battery and the status of the device.
If the operation is continued without line power, MAC 1200/ 1200 ST will shut down itself when a minimum level is reached to prevent the battery from a excessive discharge.
Functional Description: PCB Control CS_CI PCB Control CS_CI The PCB Control CS_CI is the main board of the MAC 1200 resting ECG analysis system. The main board accommodates the control functions of the device, except the line power supply and the NC battery-charging controller. Generation of Internal Power Supplies Logic Power Supply +5 V and +3.3 V The processor core and the memory is a fully +3.3 V design.
Functional Description: PCB Control CS_CI Switch On/Off Circuit The Switch On/Off circuit consists of the On/Off control and the Voltage control. The On/Off control switches the device on or off by switching the device supplies +UVERS and +USUPPSW. The On/Off control is supplied from the +5 VSTB. Device On Sequence The transition from device Off state, or Standby state when line power is connected, to device On state can only be activated by pressing the On/Off button.
Functional Description: PCB Control CS_CI ECG Recording and Front-End Processing The patient input is classified as being cardiac floating and is defibrillation-proof. The patient leads are connected with a 15-pin connector with a special form for the specified and released patient cables for the MAC 1200 resting ECG analysis system.
Functional Description: PCB Control CS_CI Controller Core The actual core comprises the Motorola Power PC MPC855T, which contains the following integrated components. Chipselect logic DRAM controller SCC and SMC for RS-232 SPI interface I/O ports ...
Functional Description: PCB Control CS_CI The Reset Configuration defines startup conditions like boot port size and clock generation source. Four LEDs indicate the device status in addition to the following. LED1: Active when a HRESET_ occurs. LED2: Indicator for the logic supply. LED3, LED4: Indicate internal software states.
Functional Description: PCB Control CS_CI Memory The complete memory of the MAC 1200 resting ECG analysis system is located on the PCB Control CS_C(I). The software of the device can be loaded through the JTAG port during the production process, or for service purposes with the appropriate programming software through the RS-232 interface.
Page 31
Functional Description: PCB Control CS_CI LCD Power Supply The LCD power supply VEE of –23 V is generated from the +5 V logic supply. The generation starts after HRESET_ becomes inactive to ensure that the logic supply of +5 V first is applied to the display. Adjusting Contrast Contrast adjustment of the display is accomplished with the contrast voltage V0.
Page 32
Functional Description: PCB Control CS_CI Keypad Interface The keypad interface contains the control register for 8 keypad columns and the receiving register for 7 keypad rows, and the control signals for the status LEDs: LED_LIN, LED_BATT_, LED_START, and LED_STOP. Using the matrix of 8 x 7, up to 56 keypads can be detected. Identification of the key pressed is as follows.
Page 33
Functional Description: PCB Control CS_CI Thermal Printhead Temperature Monitoring The thermal printhead temperature monitoring measures the temperature of the thermistor, located on the printhead. A constant current source effects a temperature-dependent voltage drop. If a printhead temperature of 55°C is exceeded, the signal REC_OVHEAT is activated and the heating of the printhead is prevented by disabling the signal STROBE.
Page 34
Functional Description: PCB Control CS_CI RS-232 Interface The MAC 1200 resting ECG analysis system has an RS-232 interface accessible with a 9-pin Sub-D connector. Except for the RS-232 driver chip, the interface is integrated into the control core of the MPC855T. The interface has the following attributes.
(2018357-003 and 2013503-003). The main board generates a supply voltage of +8 Vdc, to operate a special GE modem direct on the RS-232 interface on the MAC 1200 device without the need for an external power supply for the modem.
Functional Description: Internal Interfaces Internal Interfaces Mechanical Interfaces Mechanical interfaces are described in “Mechanical Components” on page 1-8. Electronic Interfaces This section describes the pinning, function and significance of the signals on the internal interfaces of the functional components. Interface to the Switching Power Supply The interface to the switching power supply is realized by the connector POSUP/ on the PCB Control CS_CI.
Page 37
Functional Description: Internal Interfaces BATT/ Pin Number Signal Input/Output Function Definition +24V Output Supply battery charging +24V Output Supply battery charging +24V Output Supply battery charging Code Coding Pin GND24V Common Ground GND24V From power supply GND24V After ferrite decoupling GND24V +UBATT Input...
Page 38
Functional Description: Internal Interfaces Interface to the Keypad The interface to the keypad is realized with the connector KEYB/. It contains the signals for the keypad rows and columns, the supply and the control signals for the status LEDs of the keypad. The foil connecting cable is part of the keypad itself. Connector denotation: KEYB/ Type:...
Page 39
Functional Description: Internal Interfaces Interface to the Motor The motor control signals are supplied through a 6-pin connection cable, plugged to the connector MOTOR/ on the PCB Control CS_CI. Connector denotation: MOTOR/ Type: Multipoint connector 1 x 6 pin, 180°, Reverse terminal protection, MODU II Interfaces for Production Tests Debug Port...
Page 40
Functional Description: Internal Interfaces In Circuit Programming Port With the In Circuit Programming Port, both CPLDs can be programmed or updated in a Daisy Chain queue. Connector denotation: ISP/ Type: Male connector, 2 x 5-pin, 180°, Pin 4: coding pin The function of the individual pins is given in the following table.
Functional Description: Interfaces to Peripherals Interfaces to Peripherals The MAC 1200 resting ECG analysis system has only 3 interfaces for peripherals. Mains input Patient input RS-232 interface The mains input interface on the device is a 3-pin standard cold appliance socket connection, which is integrated into the mains input module.
Functional Description: Interfaces to Peripherals Electronic Interfaces RS-232 Interface The 9-pin sub-D Connector of the RS-232 interface is implemented on the PCB Control CS_CI directly. From Version V1.1, the remote start input pin is available on pin 8. In MAC1200 US devices the modem supply output +8 V is available on pin 6 additionally.
Page 43
Functional Description: Interfaces to Peripherals Patient Input The connection to the patient cable is realized with a 15-pin female sub-D Connector, implemented on the PCB Control CS_CI directly. The mechanical plastic housing ensures that only patient cables that have been released for use with the MAC...
Functional Description: Limitations Limitations The following operating modes are not implemented in the MAC 1200 resting ECG analysis system. No SpO2 No spirometry No late potentials, no RR variability No phono, no US Doppler A scope output is not available. ...
System Test Function: General Information General Information NOTE The master password used to override all other passwords is SYSTEM. All messages displayed during the self-test are in English. The functions for the system test are mostly menu-guided. Some tests require special auxiliary resources. These include interface testers or PC, connection cables, signal generators, etc.
System Test Function: Test Start Test Start After simultaneously pressing the keys Shift + 12 Lead (or Auto), the initial display menu appears, so you can select a specific test. The specific test functions are activated by pressing the corresponding key. Table 1.
System Test Function: Display Test Display Test After pressing the 1 key, the following menu appears. Table 2. Display Test Test Performed Function Test pattern Pressing this key generates a vertical stripe pattern. Each subsequent pressing generates the inverse display of its predecessor. Clear display All display pixels are inactive.
System Test Function: Keyboard Test Keyboard Test When the 2 key is pressed, the following menu appears. Table 3. Keyboard Test Test Performed Bleeper Test When a key is pressed, this key or its function is displayed. This can either occur by a triple character display, e.g., “AAA”...
System Test Function: Motor Test Motor Test When the 3 key is pressed, the following menu appears. Table 4. Motor Speed Test Test Performed Format/Speed Speed selection motor Start/Stop Start/Stop motor Terminate with any key. The speed select key is used to set the required speed and the motor is set into motion with Start/Stop.
System Test Function: Test Results Test Results Pressing the 4 key triggers the output of the test results. The output of the test protocol supplies the following information. Time and date Device serial number Device model Software release ...
System Test Function: Self-Test Self-Test The test results outlined in “Test Results” on page 3-8 are gathered during the self- test, which is always performed on power-up. if errors are detected, a message appears on the display after the test, indicating the possible errors. The following error codes are used to identify the error.
System Test Function: Recording Test Recording Test The recording test is activated by pressing the 5 key. A window with 2 channels is displayed. Table 5. Motor Speed Test Test Performed Speed Selects the speed Start/Stop Starts and stops a real time 2 channel recording The selected speed and sensitivity are displayed.
System Test Function: V24 Tests V24 Tests Several possibilities are available to test the serial interface. Signal transmission and receiving of the interface can be tested by an internal feedback from TXD and RXD. Signal transmission and receiving can also be tested with a remote station. For the test with a remote station, the connection cable from MAC...
System Test Function: V24 Tests Send and Receive with Device (3) Sending and receiving can be tested with a remote device by pressing the 3 key. The device should be connected as described in “Send Test String to Device (2)”. After pressing the 3 key a test string is sent to the remote station.
System Test Function: Time and Date Time and Date Follow these instructions for checking and setting the time and date. Pressing the 7 key calls up the following display. Table 7. Time and Date Time (hh:mm) [18:25:29] Date (dd.mm.yyyy) [03.03.2003] To set time or date, select the corresponding time or date field by pressing the return key.
System Test Function: Electrode Test Electrode Test The electrode test is started by pressing the 8 key. The following menu is displayed. Table 8. Electrode Test Electrode Test Terminate with any key Select pace electrode (key P) Pace electrode LA Cable = 10 electrodes (or no cable) The status of each electrode is displayed.
System Test Function: Time Constant/Baseline Roll Filter Time Constant/Baseline Roll Filter Beginning with software release V5.2, the time constant can no longer be set in the service mode. It is accessible in the user mode under “System Setup”, Baseline roll filter, 0.04 Hz, 0.08 Hz, 0.16 Hz.
System Test Function: Setting the Device Model Setting the Device Model The setting of the device model is activated by pressing the D key, the following display image appears. Table 9. Device Model Europe 2 JAPAN RUSS PHARMA (US devices only) WARNING Changing the device model affects the selection of the output formats and languages available, and can switch the program for...
Page 61
System Test Function: Setting the Device Model The table below shows the most important combinations: Table 10. Model Combinations Default Settings Device Model Interpretation Output Formats Languages Override Line Filter Function Enabled 1 INT HEART international international 50 Hz 2 USA 12SL 60 Hz 3 Europe 2...
System Test Function: Interpretation (International Device Models Only) Interpretation (International Device Models Only) Switching over the program for interpretation is activated by pressing the H key, the following display image appears: Table 11. Interpretation HEART 12SL The currently enabled interpretation program appears in reverse video. WARNING Switching over to another program for Interpretation affects the measurement results and the interpretation.
System Test Function: Pace Enhance Pace Enhance Beginning with software release V5.2, the Pace Enhance mode can no longer be set in the service mode. It is accessible in the user mode under “System Setup”, Pace Enhancement, YES or NO. Revision D MAC...
System Test Function: Serial Number Serial Number The serial number of the device is displayed by pressing the N key. The serial number is only displayed and cannot be changed. The serial number is entered during the manufacturing process. ...
System Test Function: Ergometer / Treadmills (JAPAN Device Model Only) Ergometer / Treadmills (JAPAN Device Model Only) Enable or disable ergometer/treadmill functionality in the Japanese device model by pressing the “O” key. The following display image appears. Table 12. Ergometers / Treadmills YES (drivers are available in Stress Test mode) NO (not available) Revision D...
System Test Function: ACI-TIPI (US Device Models Only) ACI-TIPI (US Device Models Only) Enable or disable ACI-TIPI functionality by pressing the “O” key, the following display image appears: Table 13. ACI-TIPI ENABLE DISABLE NOTE ACI-TIPI can be DISABLED in the service mode at any time, independently of the the Patient Data Setup screen.
System Test Function: Pharma Pharma The MAC 1200 device variant is configured in the service mode and any MAC 1200 can be configured as a PHARMA device.The Pharma device variant is designed for use by Clinical Research Organizations (CROs) in pharmaceutical trials. MAC 1200 units with the Pharma option are configured for specific pharmaceutical trials by CROs and then sent out to trial sites to collect data.
Page 68
System Test Function: Pharma For your notes 3-24 MAC 1200 resting ECG analysis system Revision D 2012250-095...
Before replacing the primary fuses in the power input module, the device should also be switched off and the mains plug disconnected. For replacing components, only the original GE components, mentioned in the spare parts list may be used. ...
Repair Instructions: Replacing Components Replacing Components For all the following points, the “Safety Instructions” on page 4-3 need to be observed. WARNING The device has to be switched off. Opening the Device To open the device, follow this sequence. Release the five fastening screws on the underside of the device. Open the paper feed flap.
Page 73
Repair Instructions: Replacing Components Pull forward on lower part of connector port and small plastic tray will pop out with fuses. Remove fuses. Align and insert new fuses in plastic fuse tray. Position plastic fuse tray and snap back into position. Revision D MAC...
Page 74
Repair Instructions: Replacing Components Replacing the Keypad Open the device as described in “Opening the Device” on page 4-4. Replace with new Keypad FRU Assembly by connecting keypad connector in front of unit. Place the plastic spacer tubes back in the handle. Align upper shell to lower unit.
Page 75
Remove the battery and disconnect the battery plug BATT_IN/ from the PCB Battery Charge. WARNING Only original GE-supplied batteries may be used. (See Chapter 8, “Parts List” for details.) Defective NC-batteries or empty batteries should be disposed of in accordance with the applicable legal stipulations or returned to the factory.
Page 76
Repair Instructions: Replacing Components Replacing the PCB Control Before replacing the PCB, if still possible, print out the settings configured by the user and check the options enabled. Open the device. Disconnect the plug-in connector POSUP/ to the power supply unit, and the connector BATT_IN/ from the PCB battery charge.
Page 77
Repair Instructions: Replacing Components Replacing the Graphics Display Follow these instructions to replace the graphics display: Open the device. Open connector HOS/ and release the flat ribbon cable of the display. Undo the screws of both plastic display holders. Shift the display 2 mm in the direction to the handle and lift up display and holders.
Page 78
Repair Instructions: Replacing Components For your notes 4-10 MAC 1200 resting ECG analysis system Revision D 2012250-095...
Troubleshooting: Basic Information Basic Information Repairs may only be conducted by approved service personnel. For all the following points, see Chapter 4, “Safety Instructions” section for items that need to be observed. NOTE The master password used to override all other passwords is SYSTEM. Device cannot be switched on even though power plug is plugged in ...
Page 82
Chapter 3, “Self-Test”. The error codes refer to the PCB Control, and should be noted as information for the service center. MAC 1200 Fails to Give Printout, No Paper Transport Perform the following test in the 6-lead operating mode.
Page 83
Troubleshooting: Basic Information MAC 1200 Only Prints on the Upper or Lower Section of the Printout Is the paper feed flap is only engaged on one side? MAC 1200 Prints, but only Baselines are Printed Out Are the electrodes applied correctly? ...
Page 84
Troubleshooting: Basic Information For your notes MAC 1200 resting ECG analysis system Revision D 2012250-095...
Adjustment Instructions: Basic Information Basic Information For the MAC 1200 resting ECG analysis system, no adjustment of components is required. When PCB Control CS_CI, or the graphics display has been exchanged, the display contrast should be adjusted to an optimum contrast ratio as described in the user manual.
Page 88
Adjustment Instructions: Basic Information For your notes MAC 1200 resting ECG analysis system Revision D 2012250-095...
Service and Maintenance: Technical Inspections/Preventative Maintenance Technical Inspections/Preventative Maintenance It is recommended that a technical inspection/preventative maintenance check is performed once a year. The following checks/tests are to be performed. Check device and accessories for mechanical defects which impair their function.
Service and Maintenance: Technical Inspections/Preventative Maintenance Visual Check Device and accessories should be checked for the following: Fuse cartridges comply with vendor's specifications. Labels and inscriptions on the device relating to safety are clearly legible. The mechanical state of the device permits its further use. ...
Page 93
Service and Maintenance: Technical Inspections/Preventative Maintenance Device Test Result Check Output “Test Results” as detailed in Chapter 3, “Test Results”. Main parameters: All memory stores free from error? ASIC test O.K? Sample rate 1000? Printhead voltage >18 V, battery charge O.K? ...
Page 94
Pace setting Pace amplitude 6 mV Pace duration 0.2 ms Adjust manual (6-lead) mode on MAC 1200 device to be tested and select lead groups I, II and III. MAC 1200 resting ECG analysis system Revision D 2012250-095...
Page 95
Service and Maintenance: Technical Inspections/Preventative Maintenance Start the recording. The pace pulses must be visible as spikes on the recording output. Identification of Disconnected Electrodes Reset the simulator to ECG signal at a heart rate of 60 bpm. Remove one electrode after the other from the ECG simulator.
Service and Maintenance: Technical Inspections/Preventative Maintenance Safety Analysis Tests General Information The suggested safety analysis tests refer to the international standard IEC 601-1. The tests are generally performed with safety testers, the measuring circuits of most of which are calibrated according to IEC 601. The following is a general description of the tests to be performed.
Page 97
Service and Maintenance: Technical Inspections/Preventative Maintenance Enclosure Leakage Current Test This test is performed to measure leakage current from chassis to ground during normal condition (N.C.) and single fault conditions (S.F.C.). In all cases, the leakage current is measured from any exposed conductive parts to ground, the unit being tested has to be switched on and off.
Page 98
Service and Maintenance: Technical Inspections/Preventative Maintenance Test has failed if the measured values are greater than 50 µA. Electrical Diagram for Patient Leakage Current Test Power Cord Safety Tester High NORM Power Cord Patient Cable or Test Plug 022A For the protection of the test person, the following values of resistor R may be used. Type BF 22 kOhm (120 to 130 V)...
Service and Maintenance: Maintenance, Cleaning, and Disinfection Maintenance, Cleaning, and Disinfection MAC 1200 resting ECG analysis system maintenance, cleaning, disinfection is performed in accordance with the “Cleaning, Disinfection and Maintenance” chapter of the operator manual for the MAC 1200 resting ECG analysis system. Applications requiring extensive recordings may result in deposits on the thermal array printhead, which normally do not have any adverse effect on the printing quality.
Page 100
Service and Maintenance: Maintenance, Cleaning, and Disinfection For your notes 7-12 MAC 1200 resting ECG analysis system Revision D 2012250-095...
Introduction The field replaceable units listed in this chapter cover two versions of the product: MAC 1200 with the -002 board MAC 1200 with the -003 board Most of the FRUs listed in this chapter apply to both versions of the product. Any FRUs that are specific to one of these product configurations, have appropriate notes added to the “Description”.
MAC 1200 Europe 2 Basic unit with Battery Without Pump 10116830 MAC 1200 Europe 2 Basic unit with Battery With Pump Table 2. MAC 1200 ST Part Number Description 10116831 MAC 1200 ST International Basic unit with Battery Without Pump 10116835...
Page 105
Parts List: Field Replaceable Units (FRUs) Table 3. MAC 1200 and MAC 1200 resting ECG analysis system Manuals Part Number Description Language 2012250-021 MAC 1200 resting ECG analysis system Operator’s Manual English (Int’l) 2102250-022 MAC 1200 Operator’s Manual English, USA 2012250-038 MAC...
Page 106
2000076-001 Upper case with keypad and filter pane (USA) 2000077-001 Upper case with keypad and filter pane (Europe 2) Table 6. Upper Case for MAC 1200 ST Part Number Description 2006893-001 Upper case with keypad and filter pane (INT 1)
Page 107
43252210 Plastic holder for Printhead right side 41515439 Pressure spring for Printhead 50465725 Metal Tube 50465774 Contact Plate 50465726 Press Strip Table 8. Keypads MAC 1200 Part Number Description 39000200 Keypad International 39000201 Keypad Europe 2 39000202 Keypad USA 39000203...
Page 108
Table 11. MAC 1200 PWAs (version -002 board) Part Number Description 2000072-002 Spare PCB Control with remote start for MAC 1200 NOTE: (-002 board only) Backward compatible with all versions prior to v6.2. Not compatible with v6.2 or higher 2000072-002-1 Exchange PCB (-002 board.
Page 109
Parts List: Field Replaceable Units (FRUs) Table 12. Electronic Components / PCBs Part Number Description 91541936 Line filter with fuse holder 30344270 Battery 38803290 PCB battery charging (old board) 93011858 PCB power supply 91208449 Fuse T 1.25 A 92916717 Battery 3 V 0.255 Ah 30344291 Pump 12 V (Standard) 38803125...
Page 110
Connection cable from MAC 1200 to Modem 9-pin connector 22337802 Connection cable from MAC 1200 to Modem 25-pin connector (cable modem - CardioSmart) 22337804 Connection cable from MAC 1200 to Modem 25-pin with supply (US only) 22336604 Connection cable from MAC 1200 resting ECG analysis system to EC1200...
Resolution of the recording. Vertical: 8 dots/mm Horizontal: 25 mm at 25 mm/s Printer Paper GE CONTRAST Z-fold pad, 150 pages per pad, equivalent to a chart length of approximately 45 m. Paper width: 8.5 inch Sheet length: 11 inch To prevent damage to the printhead use only the original GE CONTRAST paper or the GE thermal paper with queue holes or marks.
Specifications: Basic Information Paper Transport Paper speed 5-25-50 mm/s, key selectable Error limits: at 25 and 50 mm/s, typ. ±1% at 5 mm/s, ±10% maximum At paper end, the recorder emits an audio signal and stops recording the last pages of the pad bear a colored stripe in the lower margin Membrane Keypad Pushbuttons with tactile feedback.
Specifications: Basic Information Detection of Pacer Pulses Pulse length between 0.1 and 2.5 ms Pacer pulse marker independent of pulse polarity Pulse amplitude between ±5 mV and ±700 mV Heart Rate Indication Derivation of the heart rate from all ECG signals. ...
Specifications: Basic Information Data Interface One serial RS232 interface for exchange of data with suitable external devices and software handshake. RS232 interface (standard V.24 interface): Input voltage range: ±15 V maximum Output voltage range ±5 V minimum Interface protected from electrostatic discharge for ±10 kV maximum. Pin Assignment of Data Port 5 circuit reference 3 TXDE...
Specifications: Basic Information Signal Transmission Patient Input to Recording After lead formation and digitization, simultaneous transmission of all electrode signals to the digital processing system; muscle filter, AC filter, pacing pulse identification, automatic or manual sensitivity adjustment, automatic baseline adjustment and drift compensation by means of the anti-drift system (A.D.S.) can be enabled or disabled simultaneously for all channels;...
Specifications: Basic Information ECG Calibration Automatic recording of a defined voltage step, valid for all channels. Calibration voltage, referred to ECG signal input: 1 mV Calibration pulse width on recording depends on paper speed 25 mm/s: 5 mm 50 mm/s: 10 mm 5 mm/s: 1 mm...
Specifications: Basic Information Blocking Rapid charge reversal of the coupling capacitors in the preamplifiers after electrode application ensures that the baseline is quickly restored to its original position after overranging. Electrode Monitoring Audible and visual indication on the LCD of disconnected electrodes or line break; each single electrode is monitored.
Specifications: Basic Information Mains Operation Instrument design in protection class I according to IEC 601-1 Rated voltage range: 95 to 240 V Operating voltage range: 85 to 264 V, 49 to 65 Hz Rated current: 0.2 ... 0.6 A Fuse: 2 x T 1.25 A, 5 x 20 ...
Specifications: Basic Information Environment Operation Temperature: between +10 and +40 °C / 50 and 104°F Relative humidity: between 25 and 95% Atmospheric pressure: between 700 and 1060 hPa (between 20.67 and 31.3 inches Hg) Transport and Storage ...
Page 122
Specifications: Basic Information For your notes 9-12 MAC 1200 resting ECG analysis system Revision D 2012250-095...
388 032 90 R, sheet 1 388 032 90 PCB Battery Charge CS_CSI NOTE See MAC 1200 Service Manual, PN 2012250-023, for drawings that apply to the MAC 1200 with the -002 PWA board. 10-2 MAC 1200 resting ECG analysis system Revision D...