5.1.2 A1 - Alarm area

This area shows the text of the enabling alarm/s and the priority / status of the same and a
lock icon indicating whether the touch screen is enabled.
CAUTION
Three different alarms can be displayed at most and simultaneously; any
further alarm signals will be displayed in sequence at intervals of 3/5
seconds.
Alarms configurable by User
Low / High Pressure of Airways
Low / High Respiratory Rate
Low / High Expiratory Volume
Low / High Volume Minute
Low / High PEEP
Low / High O
Gas Sensor: Low / High EtCO
On Battery (Power Failure - MENU
SETUP - Other)
Apnoea (MENU SETUP- Other)
WARNING !!
When this 1000-hour operation is reached, this symbol appears.
Contact the Siare Service Center or authorized Siare for
preventive maintenance.
Falco 202 Evo
Concentration
2
2
( * ) MAINTENANCE 1000 HOURS
System alarms
Low Battery: 50% Remaining
Low Battery: 25% Remaining
Low Battery
Battery Disconnected
Battery Charger Disconnected
Battery Overtemperature
(Patient) Circuit Disconnected
Low O
Supply
2
Turbine Failure
Turbine Overtemperature
Turbine Overcurrent
MAINTENANCE 1000 hours (* a
symbol replace the alarm bell image).
5-3