Standard Event Register - Keysight U2722A Programmer's Reference Manual

Usb modular source measure units
Hide thumbs Also See for U2722A:
Table of Contents

Advertisement

2
SCPI Status Registers

Standard Event Register

The Standard Event register reports the following types of instrument events:
power-on detected, command syntax errors, command execution errors, self-test
or calibration errors, query errors, or when an *OPC command is executed. All of
these conditions can be reported in the Standard Event summary bit through
enable register. To set the enable register mask, key in a decimal value to the
register using the event status enable (*ESE)command.
Bit Definitions: Standard Event Register
Bit number
0 Operation Complete
1 Not Used
2 Query Error
3 Device-Dependent Error
4 Execution Error
5 Command Error
6 Not Used
7 Power On
The Standard Event register is cleared when:
– you execute the clear status (*CLS)command
– querying the event register using the event status register (*ESR?) command
The Standard Event Enable register is cleared when you execute the *ESE 0
command.
24
Decimal value
Definition
1
All commands prior to and including *OPC have been executed.
2
Always zero.
4
The instrument tried to read the output buffer but it was empty. Or, a new
command line was received before a previous query has been read. Or,
both the input and output buffers are full.
8
A self-test or calibration error occurred. For a complete listing of the
error messages, refer to
16
An execution error occurred (an error in the –200 range has been
generated).
32
A command syntax error occurred (an error in the –100 range has been
generated).
64
Always zero.
128
Power has been turned off and on since the last time the event register
was read or cleared.
Chapter
14,
Keysight U2722A/U2723A Programmer's Reference
"Error Messages"
on page 130.