10.12 Maintaining Recorder

Wear proper personal protective equipment (e.g. medical gloves, disposable
wrist strap, etc.) against the static electricity during maintaining recorder.
Make sure the power supply of the analyzer has been cut off before
maintenance.
If the recorder finishes a print action, wait at least 20 minutes until the
recorder head is cooled completely and then maintain the recorder.
Do not add unnecessary force to the recorder head.
Wipe off the alcohol remaining on the recorder head in time after cleaning
the recorder head.
Make sure the alcohol volatilizes completely before using the recorder to
print.
Maintain the recorder strictly as instructed below.
You should maintain the recorder every two months. Do as follows to maintain your recorder:
1. Turn off the analyzer and cut off the power supply;
2. Open the recorder door and take out the recorder paper;
3. Gently wipe the roller from left to right using cotton swabs;
4. Roll the roller and repeat step 3 to clean off all debris and stains on the roller;
5. Gently wipe the heating part of the recorder head from left to right using cotton swabs
dipped with alcohol (no drops) to clean off all debris and stains on the head;
6. Wipe off the alcohol remaining on the heating part of the recorder head using dry cotton
swabs;
7. Wait at least 20 minutes till alcohol on the heating part of the recorder head volatilizes
completely, and then install recorder paper and close the recorder door.
Maintaining Your Analyzer
10-39