displayed on the screen;
4. When the analysis is finished, the result will be displayed on the screen and the sample
probe will be repositioned.
In the prediluted mode
1. At "Calibrator" screen, press [F1] to select "Prediluted" mode;
2. Press [DILUENT] and a message box will pop up to instruct you how to dispense the
diluent into the sample tube, as Figure 9-4 shows;
3. Present a clean sample tube to the sample probe and make sure the tube is tilted
towards the probe, as Figure 9-5 shows, to avoid spills and bubbles. Press the aspirate
key to dispense 0.7mL of diluent (the dispensing volume is controlled by the analyzer)
into the tube.
4. When the dispensing is done, press [ENTER] to close the message box;
5. Make sure the calibrator is at room temperature and not expired;
6. Add 20µL of calibrator into one of the prepared sample cups and mix them well;
7. Present the mixed calibrator to the sample probe;
8. Press the aspirate key to start the run. When you hear a beep and the sample probe is
out of the cup, remove the sample;
9. The sample probe will retract into the analyzer and the analysis progress will be
displayed on the screen;
Using the Calibration Programs
Figure 9-4 "Add Diluent" window
Figure 9-5 How to dispense diluent
9-7