Do you have a question about the BC-2800 and is the answer not in the manual?
Questions and answers
Claudina
May 23, 2025
Hola! Tengo un error en mi BC 2800, me aparecio error: presion baja, ya realice el auto test y ahora me dice que la presión es normal, pero no se borra el error de la pantalla principal y no puedo usar el equipo
1 comments:
Mr. Anderson
May 25, 2025
If a low pressure error persists on the Mindray BC-2800 even after an auto test indicates normal pressure, you should call Mindray Customer Service Department or the distributor.
This answer is automatically generated
Raju
April 23, 2025
WBC bubbles
1 comments:
Mr. Anderson
May 12, 2025
WBC bubbles in the Mindray BC-2800 are caused when the actual WBC count time is less than the preset WBC count time by 2 seconds. Possible causes include:
1. Insufficient diluent or rinse. 2. Loose tubing connections. 3. Inappropriate WBC count time setting.
This answer is automatically generated
Bett
February 13, 2025
Abnormal background display on the screen
1 comments:
Mr. Anderson
February 13, 2025
An abnormal background error on the Mindray BC-2800 can be caused by several factors:
1. Dirty or Contaminated System – Cleaning tests should be performed to remove any buildup or debris. 2. Electrical Issues – Poor grounding or electrical noise can interfere with the system. 3. Reagent Problems – Ensure the use of genuine Lyse reagent. 4. Electrode Issues – A greenish color on the electrode inside the counting bath may indicate a soldering problem between the silver plate and the copper wire. Resoldering or replacing the bath may be necessary. 5. Faulty Tubing or Connections – Loose tubing connections can contribute to the error. 6. Startup Procedure Needed – Running the startup procedure by pressing [F3] at the Count screen may resolve the issue.
If the issue persists after these steps, contacting Mindray Customer Service or a distributor is recommended.
This answer is automatically generated
Pradeep
January 21, 2025
A/d interrupt abnormal
1 comments:
Mr. Anderson
May 14, 2025
"A/D interrupt abnormal" on the Mindray BC-2800 indicates a problem with the analog-to-digital (A/D) interrupt function. This may suggest an issue with the system’s ability to properly convert analog signals to digital data, which could affect test accuracy. The test for this function is done by selecting “A/D interrupt” in the “Circuit” group. If the test result is not normal, service may be required.
BC-2800 Auto Hematology Analyzer Service Manual...
Page 3
Mindray equipment. No part of this can be disseminated for other purposes. In the event of inadvertent or deliberate publication, Mindray intends to enforce its rights to this work under copyright laws as a published work. Those having access to this work may not copy, use, or disclose the information in this work unless expressly authorized by Mindray to do so.
Page 4
This equipment must be operated by skilled/trained clinical personnel. Warning It is important for the hospital or organization that employs this equipment to carry out a reasonable maintenance schedule. Neglect of this may result in machine breakdown or injury of human health. BC-2800 Auto Hematology Analyzer...
Page 5
Safety, Reliability and Performance Mindray is not responsible for the effects on safety, reliability and performance of the BC-2800 Hematology Analyzer if: ■ assembly operations, extensions, re-adjusts, modifications or repairs are carried out by persons other than those authorized by Mindray.
Copyright Return Policy Return Procedure In the event that it becomes necessary to return a unit to Mindray, the following procedure should be followed: Obtain return authorization. Contact the Mindray Service Department and obtain a Customer Service Authorization (Mindray) number. The Mindray number must appear on the outside of the shipping container.
To monitor such system status as the +48V, +12V and -12V supplies of the analog board, the +3.3V and +12V supplies of the CPU board itself and the temperature of the whole analyzer. To receive the keypad signal and control the keypad buzzer and LCD backlight. BC-2800 Auto Hematology Analyzer (V1.0) 1- 1...
Page 10
To generate control signals to control the valves, aperture zapping, HGB LED, current source and digital pot. To drive and turn on the LCD and adjust the contrast. To drive the keyboard, printer and floppy drive. BC-2800 Auto Hematology Analyzer (V1.0)
+3.3V supply is also further converted into a +1.5V supply to power the FPGA. The +12.8V supply serves the CPU board only. Figure 1-2 Power distribution of the CPU board BC-2800 Auto Hematology Analyzer (V1.0) 1- 3...
Page 12
Figure 1-3 Arrangement of the CPU Clock The X1, X4 and X2 are external crystal oscillators whose frequencies are 45MHz, 45MHz and 24MHz respectively. The clock output of the CPU, BCLKO, is main clock signal of the CPU board. BC-2800 Auto Hematology Analyzer (V1.0)
The Watch-Dog-Timer (WDT) is TI TPS3828. It monitors the running of the software. The CPU must send a feedback to the WDT every 1.6s, otherwise the WDT will force the CPU to restart. BC-2800 Auto Hematology Analyzer (V1.0) 1- 5...
Page 14
CPU on-line. 1.4.8 LEDs When D1 is on, it means +3.3V is functioning properly. When D9 is on, it means +12.8V is functioning properly. When D5 is on, it means the system is reading BC-2800 Auto Hematology Analyzer (V1.0)
Page 15
DOM. When D7 is on, it means the FPGA has been configured and is functioning properly. When D20 is on, it means the FPGA is restarting; The D11~D18 indicate the system status as defined by the software. BC-2800 Auto Hematology Analyzer (V1.0) 1- 7...
1.5.3 Signals of LCD Contrast The Super I/O generates PWM signals that are then integrated to output a 0~2.5V analog signal to control the LCD contrast. The user can adjust the contrast through the software interface. BC-2800 Auto Hematology Analyzer (V1.0)
PS/2). The software will automatically adapt to the connected printer or the floppy drive. The Super I/O provides a keyboard interface and a mouse interface (COM3 and COM4). Note that the BC-2800 does not support the mouse yet. 1.6.3 GPIOs 1 Signals of the Start key The FPGA detects the input signal, which will turn low when the Start key is pressed.
Page 18
5 Signals controlling valves and pumps The Super I/O outputs 20 control signals to control the valves and pumps through the driving board. Since the BC-2800 only has 1 pump and 11 valves, the redundant lines and ports are reserved.
Figure 1-8 Switching control circuit 1. Photocoupling circuit The photocoupling circuit mainly consists of the photocoupler and resistors. It provides 16 TTL outputs to the valves and pumps. The photocoupler, TLP521-2, BC-2800 Auto Hematology Analyzer (V1.0) 1- 11...
Page 20
The driving voltage of the valves and pumps are 12V (TTL). The circuit mainly consists of ULN2068. In the BC-2800, the circuit can maximum drives 14 valves and 2 pumps. The fluidic system decides which pumps or valves are to be actually used.
Page 21
The testable signals are: control signals of valves and pumps, sequence signals of the motors, valves, position signals sent by the position transducers, serial communication signals, reset signals and voltage signals of the power supplies. BC-2800 Auto Hematology Analyzer (V1.0) 1- 13...
The adapter incorporates two FPC/FFC connectors, J2 and J3. The J3 is for the BC-2800 display while the J2 is reserved for other Mindray analyzer. Only the J2 is installed for the BC-2800. The J1 serves to connect the LCD signal cable.
Page 23
Hardware Figure 1-10 Schematic of the adapter BC-2800 Auto Hematology Analyzer (V1.0) 1- 15...
The backlight board uses an independent 12V power supply and receives the control signals through photocouplers. The transistor is used to help control the LED so that the power indicator can be turned on even when something is wrong with the MCU. 1-16 BC-2800 Auto Hematology Analyzer (V1.0)
Page 25
The buzzer is controlled by a DC signal (5V DC; current<40mA). The 5V supply of the buzzer is isolated from the VDD and the control signal is received through a photocoupler (TLP521-2)that is controlled by a current around 10mA. Figure 1-12 Buzzer control BC-2800 Auto Hematology Analyzer (V1.0) 1- 17...
Chapter 2 Fluidic System 2.1 Fluidic system prepares diluent for the whole blood and prediluted modes counts blood cells and measures HGB concentration dispenses diluent implements the flush, startup and cleaning cycles controls vacuum/pressure BC-2800 Auto Hematology Analyzer (V1.0) 2- 1...
Sleeve Left and right photo-couplers Retaining ring Figure 2-1 Sample probe assembly Mount 2.2.2 Syringe assembly 2.2.3 Pump Assembly 2.2.4 Volumetric metering unit 2.2.5 Right valve panel and front panel See Chapter 3 Disassembling Instructions. BC-2800 Auto Hematology Analyzer (V1.0)
The solenoid valve is ASCO458. Totally 11 valves are used, 7 three-way valves and 4 two-way valves. The two syringes, 10mL and 50uL, are made by Mindray. The aperture is Ø80. The sample probe has two layers and its surface is polished, making the probe easy to clean.
The aspiration/dispensing module includes a motor that drives the 50 uL and 10 mL syringes, the former for aspirating/dispensing the whole blood sample and the latter for aspirating/dispensing the prediluted sample, diluent and lyse. See the figure below for details. BC-2800 Auto Hematology Analyzer (V1.0)
Page 31
Service tip: Note that the length and type of the buffer tubing shall not be changed. 2.4.2 Counting module The counting module mainly consists of the bath, valve 1 and 2, vacuum filter, metering tube, negative pressure and other supporting components, as figure below shows. BC-2800 Auto Hematology Analyzer (V1.0) 2- 5...
Page 32
By monitoring the counting time, the system can easily know whether the aperture is clean or blocked and feed this information to the service personnel in terms of the aperture voltage so that they can service the BC-2800 Auto Hematology Analyzer (V1.0)
Page 33
NO end of valve 6.Then the plunger moves upward to dispense the aspirated diluent, through valves 5 and 6 and the buffer tubing, to the sample probe to wash the interior. BC-2800 Auto Hematology Analyzer (V1.0) 2- 7...
Page 34
The front bath is washed by diluent and the back bath by rinse. The diluent and the rinse are different and cannot replace each other. BC-2800 Auto Hematology Analyzer (V1.0)
Page 35
1 and the pump and the NC end of valve 2 to establish the vacuum. The pressure value is monitored by the pressure sensor. The pump is an imported American product, whose P/N is 3001-10-07252. BC-2800 Auto Hematology Analyzer (V1.0) 2- 9...
Page 36
During the mixing process, the airway should be well drained, or the trapped liquid will affect the quantity of the bubbles as well as the dilution. 2-10 BC-2800 Auto Hematology Analyzer (V1.0)
Page 37
Likewise, the waste of the probe wipe is discharged to the outside when valves 9 and 10 are open. Bath Negative pressure assembly chamber Waste exit Pump Probe Wipe Figure2-9 Waste discharging module BC-2800 Auto Hematology Analyzer (V1.0) 2- 11...
When analyzing a prediluted sample, the operator should first collect 20μL of capillary sample and dispense 1.6mL of diluent from this analyzer to predilute the sample. Then the operator should present the prediluted sample to the analyzer, 2-12 BC-2800 Auto Hematology Analyzer (V1.0)
Page 39
Figure 2-11 Diluting procedure of prediluted mode 2.5.1.3 Volume range of blood cells After reacting with the diluent and lyse, the cell volumes mainly fall into the following ranges: WBC: 30~350 RBC: 25~250 PLT: 2~30 BC-2800 Auto Hematology Analyzer (V1.0) 2- 13...
Page 42
Left view: 1 --- RS232 Port1 (for the scanner) 2 --- Parallel Port 3 --- RS232 Port 2 (for the computer) 4 --- Power Interface of Floppy Disk Drive 5 --- Keyboard Interface Figure3-3 Left view BC-2800 Auto Hematology Analyzer (V1.0)
Figure3-4 Removing the left, right and top covers 3.2.2 Removing the back plate and the fixing plate of the power module: As the figure below shows, remove the retaining screws (totally 10) to remove the plates. BC-2800 Auto Hematology Analyzer (V1.0) 3- 3...
Page 44
Disassembling the Analyzer Figure 3-5 3.2.3 Removing the front cover As the figure below shows, remove the retaining screws (totally 6) to remove the cover. Figure3-6 BC-2800 Auto Hematology Analyzer (V1.0)
Page 45
As the figure below shows, remove the retaining screws (totally 7) to remove the keypad. Figure3-8 3.2.6 Removing the LCD and adaptor As the figure below shows, remove the retaining screws (4 for each part) to remove the LCD and adaptor. BC-2800 Auto Hematology Analyzer (V1.0) 3- 5...
Page 46
As the figure below shows, remove the retaining screws (totally 3) to remove the box. Figure3-10 3.2.8Removing the driving board, analog board, CPU board and power board As the figure below shows, remove the retaining screws (totally 18) to remove the boards. BC-2800 Auto Hematology Analyzer (V1.0)
Page 47
As the figure below shows, remove the retaining screws (totally 2) with a socket wrench. Figure3-12 3.2.10 Removing valves As the figure below shows, remove the retaining screws (two for each valve) to remove the valves. BC-2800 Auto Hematology Analyzer (V1.0) 3- 7...
Page 48
Disassembling the Analyzer Figure3-13 3.2.11 Removing the sample probe assembly: As the figure below shows, remove the retaining screws (totally 3) to remove the assembly. Figure3-14 BC-2800 Auto Hematology Analyzer (V1.0)
Page 49
As the figure below shows, remove the retaining screws (totally 4) to remove the assembly. Figure3-15 3.2.13 Removing the vacuum chamber and volumetric metering tube assembly As the figure below shows, remove the retaining screws (4 for each part) to remove the assembly. BC-2800 Auto Hematology Analyzer (V1.0) 3- 9...
Page 50
Disassembling the Analyzer Figure3-16 3.2.14 Removing the pump assembly As the figure below shows, remove the retaining screws (totally 7) to remove the assembly. Figure3-17 3-10 BC-2800 Auto Hematology Analyzer (V1.0)
Page 51
Access the Setup→Password screen and enter the administrator password to obtain the administrator authority. Access the Service→Self-test screen. Press [F1] to select the Machine group and press [↑] or [↓] to move the cursor to Elevator motor, as the figure below shows. BC-2800 Auto Hematology Analyzer (V1.0) 3- 11...
Page 52
Press [ENTER] and an elevator motor screen will pop up, as the figure below shows. Figure3-20 Elevator motor screen Press [↑] to move the sample probe upward and press [→] to move the probe to the top of the bath, as the figure below shows. 3-12 BC-2800 Auto Hematology Analyzer (V1.0)
Page 53
Loose the retaining screw by a Philips screwdriver, as the figure below shows. Figure 3-22 Removing screws Remove the probe from the wipe block and insert the localizer into the wipe block from the bottom, as the figure below shows. BC-2800 Auto Hematology Analyzer (V1.0) 3- 13...
Page 54
Warning The probe is sharp and may contain biohazard materials, including controls and calibrators. Avoid any unnecessary contact with the probe and probe area. Follow the steps given below to replace the wipe block: 3-14 BC-2800 Auto Hematology Analyzer (V1.0)
Page 55
When the connection is done, place the wipe block back to its original position. 3. Follow steps 7~9 according to Section 3.3.1 to fix the sample probe. Figure3-25 Installing probe wipe BC-2800 Auto Hematology Analyzer (V1.0) 3- 15...
4. A histogram that does not show differential result because the peak is in the middle of the histogram. Figure 4-4 5. A histogram that does not show differential result because of increased nucleated erythrocytes, or interference or inadequate hemolysis BC-2800 Auto Hematology Analyzer (V1.0) 4- 1...
Page 58
(check the pulse graph to determine whether there is interference) in the WBC channel. Figure 4-6 7. No lyse reagent or poor hemolysis Figure 4-7 8. Increased granulocytes Figure 4-8 9. Increased lymphocytes Figure 4-9 Tumor patient Figure 4-10 Increased monocytes Figure 4-11 BC-2800 Auto Hematology Analyzer (V1.0)
Length of normal level controls data should be: WBC:400 ~ 700K RBC:250 ~ 450K PLT:300 ~ 600K Normal pulse graphs Figure 4-12 WBC pulses Figure 4-13 PLT pulses Figure 4-14 RBC pulses Figure 4-15 Normal WBC background pulse BC-2800 Auto Hematology Analyzer (V1.0) 4- 3...
Page 60
Histograms and Pulse Graphs Figure 4-16 Figure 4-17 Pulse of normal RBC background check Pulse of normal PLT background check BC-2800 Auto Hematology Analyzer (V1.0)
Page 61
(background) obviously (normal sample) Figure 4-22 Severe interference in P Figure 4-23 Severe interference in PLT LT channel and data length increases channel and data length increases obviously (background) obviously (normal sample) BC-2800 Auto Hematology Analyzer (V1.0) 4- 5...
Page 62
(normal sample) obviously (normal sample) Figure 4-28 Slight interference in RBC Figure 4-29 Inadequate or no hemolysis in channel and data length does not increase WBC channel and data length increases obviously (normal sample) BC-2800 Auto Hematology Analyzer (V1.0)
Page 63
Figure 4-33 Measuring interference from caused by inverter. Feature: sine wave inverter with cycle of 20~26us Figure 4-34 Insufficient liquid in WBC bath during count Figure 4-35 Insufficient liquid in RBC bath during count BC-2800 Auto Hematology Analyzer (V1.0) 4- 7...
Page 64
Figure 4-38 Interference in WBC channel from tubing. Feature: data length Figure 4-39 Interference in RBC channel increases, the base line of signal is not from tubing. Feature: data length stable. increases, the base line of signal is not stable. BC-2800 Auto Hematology Analyzer (V1.0)
Replacement of the aperture(s) or gasket (s) will affect WBC(whole blood), WBC (prediluted), RBC,PLT Enter the Password screen and enter the password – 3210. Press [MENU] to enter the system menu, move the cursor to Setup → Gain and press [ENTER] to enter the Gain screen. BC-2800 Auto Hematology Analyzer (V1.0)
Page 66
WBC(whole blood):gain of the WBC channel in the whole blood mode WBC(prediluted) :gain of the WBC channel in the prediluted mode RBC:gain of the RBC channel HGB:gain of the HGB gain PLT:gain of the PLT channel. BC-2800 Auto Hematology Analyzer (V1.0)
Once it occurs, immediately wipe off the spills. If it occurs inside this analyzer, be sure to shut down the power immediately and call Mindray Customer Services Department or the distributor. Otherwise, the service life of this analyzer may be shortened.
In case you spill the reagents on you skin, be sure to wash them off with much water. In case you spill the reagents into your eyes, be sure to immediately wash your eyes with much water and go see a doctor for further treatment. BC-2800 Auto Hematology Analyzer (V1.0)
Page 69
When this analyzer gives alarms for clogging, you can perform the Flush aperture or Zap aperture procedure, or press [F2] at the Count screen to unclog the system. If you see other error messages, see Chapter 7 Troubleshooting, for solutions. BC-2800 Auto Hematology Analyzer (V1.0)
Figure 6-1 Entering maintenance screen Press [ENTER] to enter the Maintenance screen, as Figure 6-2 shows. Figure 6-2 Maintenance screen If you want to exit this screen, press [MENU] to enter the system menu and access the BC-2800 Auto Hematology Analyzer (V1.0)
Page 71
Press [ENTER] to prime the tubing with diluent and the priming progress will be displayed at the bottom of the screen, as Figure 6-3 shows. When the priming is done, the screen will return to the initial state. Figure 6-3 Diluent priming screen BC-2800 Auto Hematology Analyzer (V1.0)
Page 72
Press [ENTER] to prime the tubing with rinse and the priming progress will be displayed at the bottom of the screen, as Figure 6-4 shows. When the priming is done, the screen will return to the initial state. Figure 6-4 Rinse Priming screen BC-2800 Auto Hematology Analyzer (V1.0)
Page 73
When the priming is done, the screen will return to the initial state. Figure 6-5 Lyse priming screen 6.2.4 Zap Aperture You can perform the Zap aperture procedure to unclog or prevent clogging. Follow the steps given below to do so: BC-2800 Auto Hematology Analyzer (V1.0)
Page 74
Press [ENTER] to flush the aperture and the flushing progress will be displayed at the bottom of the screen, as Figure 6-7 shows. When the flushing is done, the screen will return to the initial state. BC-2800 Auto Hematology Analyzer (V1.0)
Page 75
The soaking process will last about 5 minutes and you may press [ENTER] to stop the process any time. Note that a shortened soaking process may not be as effective as a complete one. BC-2800 Auto Hematology Analyzer (V1.0)
Page 76
Present the cleanser to the probe and press [ENTER] to aspirate the cleanser. Remove the cleanser after the probe has risen up. This analyzer will automatically soak the bath and tubing with the aspirated cleanser. BC-2800 Auto Hematology Analyzer (V1.0) 6-10...
Page 77
Unscrew and remove the retaining screws with hands or screwdrivers (pointed by the arrows shown in Figure 6-9) on the right plate. Figure6-9 Removing the two screws Follow the arrow shown in Figure 6-10 to push and remove the right plate. BC-2800 Auto Hematology Analyzer (V1.0) 6-11...
Page 78
Figure 6-10 Removing right plate Remove the screws fixing the shielding box of the bath, as Figure 6-11 shows. Figure 6-11 Shielding box Remove the shielding box to expose the bath, as Figure 6-12 shows. BC-2800 Auto Hematology Analyzer (V1.0) 6-12...
Page 79
If the lyse is still enough and the tubing is well connected to the analyzer, contact the Mindray Customer Service Department for repair.
Page 80
Check the bath and its lower hose for residual liquid. If there is, turn off the analyzer and call the Mindray Customer Service Department or the distributor for repair; if not, press [ENTER] and the analyzer will prime the bath with diluent.
Page 81
When the accumulated analyzed samples reach 4,000, a dialog box will pop up to remind to clean the wipe block. You may move the cursor to Yes and press [ENTER] to start the cleaning, or to No and press [ENTER] to ignore the message. BC-2800 Auto Hematology Analyzer (V1.0) 6-15...
Page 82
[ENTER] to drain the tubing again. Turn off the analyzer when the screen displays Turn off the analyzer. Wipe this analyzer dry and wrap it up for storage. BC-2800 Auto Hematology Analyzer (V1.0) 6-16...
Press [MENU] to enter the system menu and press the appropriate arrow keys to move the cursor to Service → Status, as Figure 6-16 shows. Figure 6-16 Entering status screen Press [ENTER] to enter the Status screen, as Figure 6-17 shows. Figure6-17 Status screen BC-2800 Auto Hematology Analyzer (V1.0) 6-17...
Service Self-test, as Figure 6-18 shows. Figure 6-18 Entering self-test screen Press [ENTER] to enter the Self-test screen, as Figure 6-19 shows. If you want acquire help information, press [HELP]. BC-2800 Auto Hematology Analyzer (V1.0) 6-18...
Page 85
Aperture(v) It measures the voltage (v) over the aperture. Vacuum It checks whether the vacuum system functions normally. Pressure It checks whether the system flushes the aperture at a normal pressure. BC-2800 Auto Hematology Analyzer (V1.0) 6-19...
Page 86
This test checks whether the recorder or printer functions normally. If normal, when you press [ENTER], the recorder or printer will print out test page; if abnormal, the screen will display the corresponding error message and you can BC-2800 Auto Hematology Analyzer (V1.0) 6-20...
Page 87
At the Self-test screen, press [F1] to select the Circuit group, as Figure 6-22 shows. You can test the A/D interrupt at this screen by pressing [ENTER] to see whether the WBC, RBC and PLT signals can be properly converted into digital signals. BC-2800 Auto Hematology Analyzer (V1.0) 6-21...
Page 89
Press [MENU] to enter the system menu and press the appropriate arrow keys ([↑][↓][←][→]) to move the cursor to Service→Log, as Figure 6-23 shows. Figure 6-23 Entering log Press [ENTER] to enter the Log screen, as Figure 6-24 shows. Figure 6-24 Log screen BC-2800 Auto Hematology Analyzer (V1.0) 6-23...
Page 90
Information column displays extra information regarding the event. This analyzer can save maximum 1000 log files and once the maximum number has been reached, the newest log will automatically cover the oldest one. BC-2800 Auto Hematology Analyzer (V1.0) 6-24...
[PgDn] to go to the previous or next screen. If you want to print out the configuration, press [PRINT]. If you want to acquire help, press [HELP]. BC-2800 Auto Hematology Analyzer (V1.0) 6-25...
Once the system finds the error has been removed, it will resume printing and process the tasks from the first one. Note that you cannot change the sequence of the queued tasks. BC-2800 Auto Hematology Analyzer (V1.0) 6-26...
Page 93
You can perform the following operations at the Print screen: Press [↑] or [↓] to select the desired task. Press [DEL] to delete the selected task. Press [HELP] to display the help information. Press [MENU] to return to the system menu. BC-2800 Auto Hematology Analyzer (V1.0) 6-27...
The chapter deals with the codes, possible causes and solutions of the errors. If the error remains after you have tried the recommended method, call Mindray Customer Service Department or the distributor. This chapter consists of two parts, the first part dealing with the errors and assigned error codes and the second possible causes and recommended solutions.
Access Setup → Password to gain the administrator authority. Access Setup → Settings → Gain to adjust the HGB blank voltage to 3.4~4.8V (4.5V recommended). If the error still remains, turn off this analyzer and call Mindray Customer Service Department or the distributor.
Page 98
→ Count and adjust the WBC count time. If the error still remains, turn off this analyzer and call Mindray Customer Service Department or the distributor. 7.2.7 RBC Clog This error message occurs when the actual RBC count time is greater than the...
Page 99
If the error still remains, access Setup → Password to gain the administrator authority and then access Setup → Settings → Count and adjust the RBC count time. If the error still remains, turn off this analyzer and call Mindray Customer Service Department or the distributor. BC-2800 Auto Hematology Analyzer (V1.0)
Page 100
Possible causes: recording paper has run out or is not correctly installed. Solution: Check whether the recording paper has run out. If so, load new paper; if not, re-install the existing paper. If the error still remains, call Mindray Customer Service Department or the distributor. 7.2.16 Recorder Too Hot Possible causes: the recording head overheats.
Page 101
Access Service → Self-test → Tubing and test the filter. The error is cleared if the result is normal. If not, call Mindray Customer Service Department or the distributor. 7.2.23 Real-Time Clock Error Something is wrong with the clock. Access Setup →...
Page 102
Something is wrong with the A/D converter on the CPU board. Solution: Access Service → Self-test → Circuit. Test the A/D interrupt. The error will be cleared if the test result is normal. Otherwise, call Mindray Customer Service Department or the distributor. 7.2.28 Vacuum Error The system does not reach the expected vacuum within the given time.
Page 103
Empty the container, or change a new container to receive the waste and re-set the waste container volume. 7.2.33 File Error Something is wrong with file saving. Turn off the analyzer and call Mindray Customer Service Department or the distributor. 7.2.34 Dynamic Memory Error Something is wrong with the system memory.
Page 105
2800-20-28847 Nut of 50µl syringe 2800-20-28848 Bolt of 50µl syringe 2800-20-28849 Connection wire of inverter 2800-21-28728 Shielding box cover assembly 2800-21-28788 Shielding box assembly 2800-30-28650 CPU board 2800-30-28652 Analog board 2800-30-28654 Driving board 2800-30-28664 Keypad BC-2800 Auto Hematology Analyzer (V1.0)
Need help?
Do you have a question about the BC-2800 and is the answer not in the manual?
Questions and answers
Hola! Tengo un error en mi BC 2800, me aparecio error: presion baja, ya realice el auto test y ahora me dice que la presión es normal, pero no se borra el error de la pantalla principal y no puedo usar el equipo
If a low pressure error persists on the Mindray BC-2800 even after an auto test indicates normal pressure, you should call Mindray Customer Service Department or the distributor.
This answer is automatically generated
WBC bubbles
WBC bubbles in the Mindray BC-2800 are caused when the actual WBC count time is less than the preset WBC count time by 2 seconds. Possible causes include:
1. Insufficient diluent or rinse.
2. Loose tubing connections.
3. Inappropriate WBC count time setting.
This answer is automatically generated
Abnormal background display on the screen
An abnormal background error on the Mindray BC-2800 can be caused by several factors:
1. Dirty or Contaminated System – Cleaning tests should be performed to remove any buildup or debris.
2. Electrical Issues – Poor grounding or electrical noise can interfere with the system.
3. Reagent Problems – Ensure the use of genuine Lyse reagent.
4. Electrode Issues – A greenish color on the electrode inside the counting bath may indicate a soldering problem between the silver plate and the copper wire. Resoldering or replacing the bath may be necessary.
5. Faulty Tubing or Connections – Loose tubing connections can contribute to the error.
6. Startup Procedure Needed – Running the startup procedure by pressing [F3] at the Count screen may resolve the issue.
If the issue persists after these steps, contacting Mindray Customer Service or a distributor is recommended.
This answer is automatically generated
A/d interrupt abnormal
"A/D interrupt abnormal" on the Mindray BC-2800 indicates a problem with the analog-to-digital (A/D) interrupt function. This may suggest an issue with the system’s ability to properly convert analog signals to digital data, which could affect test accuracy. The test for this function is done by selecting “A/D interrupt” in the “Circuit” group. If the test result is not normal, service may be required.
This answer is automatically generated