Mindray BC-2800 Operator's Manual page 175

Auto hematology analyzer
Hide thumbs Also See for BC-2800:
Table of Contents

Advertisement

The X-B graph can be interpreted as follows:
The x-coordinate represents the number of X-B analyses performed; the y-coordinate
represents the results of the X-B analyses;
For every parameter, its X-B graph can display a maximum of 500 points, 30 points per
screen. The time at which the sample was analyzed is displayed to the right of "Time". The
current cursor position and the number of all the saved points are displayed to the right of
"Loc./Total".
For every parameter, the upper dash line represents the expected result + limit;
For every parameter, the upper dash line represents the expected result – limit;
For every parameter (e.g. MCV), the three numbers to the left of the X-B graph are
defined as follows:
95.0 – expected result + limit;
90.0 – expected result;
85.0 – expected result – limit.
The "■"points fallen between the upper and lower dash lines are within the expected ranges;
The "■"points fallen outside the upper and lower dash lines are out of the expected ranges
If you see any points fallen outside the control range, do the following steps until the problem
is solved. If all the steps have failed, contact Mindray customer service department or your
local distributor for assistance.
1. Check the lower left corner of the screen for error messages. Refer to Chapter 11
Troubleshooting Your Analyzer for solutions to any displayed error messages;
2. Check the X-B settings for inappropriate entries;
3. Do the background check. In case of an abnormal background result, refer to Chapter 11
Using the QC Programs
Figure 8-25 X-B graph screen
8-19