You should perform the "Lyse prime" procedure to prime the lyse tubing when
there are bubbles in the tubing; or
the lyse in the tubing is contaminated; or
you have installed a new container of lyse without shutting the analyzer.
At the "Maintenance" screen, SELECT "Lyse prime" to prime the tubing and the priming
progress will be displayed at the bottom of the screen, as Figure 10-5 shows. When the
priming is done, the screen will return to the initial state.
10.3.4 Zap Aperture
You can perform the "Zap aperture" procedure to unclog the apertures or prevent clogging.
At the "Maintenance" screen, SELECT "Zap aperture" to zap the apertures and the zapping
progress will be displayed at the bottom of the screen, as Figure 10-6 shows. When the
zapping is done, the screen will return to the initial state.
Maintaining Your Analyzer
Figure 10-5 Priming lyse screen
10-7