Errors Without Available Error Messages - Mindray BC-2800 Operator's Manual

Auto hematology analyzer
Hide thumbs Also See for BC-2800:
Table of Contents

Advertisement

11.2 Errors without available error messages

Error
The analyzer
1. power cord is broken or not
cannot be
well connected;
turned on.
2. The fuse is broken;
3. The power outlet has no
electricity.
Liquid
drips
Damaged pump hose or blocked
from analyzer
filter.
inside.
Recorder does
1. Recorder paper is jammed;
not work.
2. Something is wrong with the
circuit.
Troubleshooting Your Analyzer
Possible cause(s)
Recommend action
1. check the power cord connection;
2. Check the fuse;
3. Check the electrical outlet.
1. Turn off the power and wipe the
analyzer dry;
2. Call Mindray customer service
department or your local distributor
for assistance.
1. Remove the jammed paper;
2. If the problem remains, turn off the
analyzer and turn it on again in 10
seconds.
11-2

Advertisement

Table of Contents
loading

Table of Contents