Operating Your Analyzer

6.3 Power-on

Place the power switch at the back of the analyzer in the ON position (1) to turn on the
analyzer. The power indicator light will be illuminated and the screen will display
"Initializing...".
The analyzer will sequentially initialize the file, hardware and fluidic systems and the whole
initializing process lasts 4 to 7 minutes, depending on how the analyzer was previously shut
down.
If any error occurs during the initialization, the analyzer will display the error messages in the
lower left corner of the screen. You should remove all the errors before running any sample.
See Chapter 11 Troubleshooting Your Analyzer for solutions.
Running samples with the abnormal background error present will lead to
unreliable results.