Advertisement

Whole Blood Mode
1. Be sure the System Status area displays "Ready";
2. Press [F2] to select the whole blood mode;
3. Present a vial of control to the sample probe so that the tip is well into the vial, and press
the aspirate key. The System Status area will display "Running" and the analyzer will
start aspirating sample;
4. When you hear a beep and the sample probe is out of the vial, remove the vial. The
sample probe will retract into the analyzer and the analysis progress will be displayed on
the screen;
5. When the analysis is finished, the result will be displayed on the screen and the
"NO./Total" in the upper left corner of the screen will automatically increase by 1 and the
sample probe will be repositioned.
If the analyzer detects WBC/RBC clogging or bubbles during the analysis,
the corresponding error messages will be displayed in the lower left corner
of the screen and the results of all the related parameters will be invalidated.
See Chapter 11 Troubleshooting Your Analyzer for solutions.
If the ambient temperature is outside the specified operating range, the
analyzer will alarm you for abnormal ambient temperature and the analysis
results may be unreliable. See Chapter 11 Troubleshooting You Analyzer for
solutions.
Deleting the QC Result
To delete the current result, press [DEL] and a message box will pop up, as Figure 8-9 shows.
CLCIK "Yes" to confirm the deletion.
Printing QC results
Press [PRINT] to print out the current QC result by the printer.
Using the QC Programs
Figure 8-9 Deleting current result
8-7