Mindray BC-2800 Operator's Manual page 262

Auto hematology analyzer
Hide thumbs Also See for BC-2800:
Table of Contents

Advertisement

C.4
Notes
This equipment must be operated by skilled/trained
Operate your analyzer strictly as instructed in this manual.
This analyzer adopts a fixed decimal point. You can enter the digits without
bothering to look for the [.] on the external keyboard.
The purpose of this analyzer is to identify the normal patient, with all normal
system-generated parameters, and to flag or identify patient results that
require additional studies.
Before connecting the power cord, make sure the power switch at the back
of the analyzer is placed in the off (O) position.
If the ambient temperature is outside the specified operating range, the
analyzer will alarm you for abnormal ambient temperature and the analysis
results may be unreliable. See Chapter 11 Troubleshooting Your Analyzer for
solutions.
Retain the shipping carton and all the packing materials, as they can be
used for packaging if analyzer must be reshipped.
Use the specified reagents.
Never use expired reagents.
To prevent contamination, tighten the container caps when the installation is
finished.
Use the printer and scanner of the specified model.
Store and use the reagents as directed by instructions for use of the
reagents.
When you have changed the diluent, rinse or lyse, run a background to see
if the results meet the requirement.
Pay attention to the expiration dates and open-container stability days of all
the reagents. Never use expired reagents.
After installing new reagents, let the reagents stand for a while before using
them.
The recorder paper is treated on one side for printing. To determine which
side is the printing side, gently scratch both sides with a fingernail and the
one with visible nail trace left is the printing side.
For the whole blood samples to be used for WBC differential or PLT count,
you shall store them at the room temperature and run them within 4 hours
Appendices