number, expiration date, open-vial stability days, expected results and limits.
The entered expiration date should be either the expiration date printed on
the labeling or the open-vial expiration date. It is earlier.
The open-vial expiration date is calculated as follows: the date that vial is
opened + the open-vial stability days.
At the "QC Edit" screen, if you want to correct an erroneous entry, MODIFY
the wrong digit.
All of the measured parameters must be calibrated before readings of this
analyzer can be used as valid analysis results.
Use the Mindray- specified calibrator. Using calibrator other than the
specified will lead to misleading results.
Refer to the instructions of use of the calibrator for how to store and use the
calibrator.
Keep dust from the prepared diluent.
Do the "Drain tubing" procedure before relocating the analyzer.
This chapter is not a complete service manual and is limited to problems
that are readily diagnosed and/or corrected by the user of the analyzer. If the
recommended solution fails to solve the problem, contact Mindray or your
local distributor.
When installing the analyzer, our authorized personnel will select a proper
communication protocol that matches the data management software
configured.
To adjust the communication protocol of the analyzer, please contact our
customer service department.
Remove the protective paper between the recorder head and the roller
inside the recorder before installing recorder paper.
Maintain the recorder strictly as instructed below.
To ensure the print quality and operating life of the recorder, it is
recommended that you purchase the recorder paper from the analyzer
manufacturer.