Maintaining Your Analyzer
Figure 10-20 Clean bath

10.3.10 Drain Bath

When at three or more of the WBC, RBC, PLT and HGB results are abnormal, you may do
the "Drain bath" procedure to find out the reason.
Follow the steps below to do so:
1. Do steps 1 to 4 of the "Drain bath" procedure to expose the bath;
2. SELECT "Drain bath" to drain the bath;
Figure 10-21 Draining the bath
3. When the draining is done, the screen will display the residual time of draining bath;