Using the Calibration Programs
Figure 9-12 Editing reference value
ENTER the reference value into the "Mean" edit box. To correct any erroneous entry,
MODIFY the digit. After you have finished the editing, press [F2] to exit the editing state.
Running the sample
After you have finished editing the calibration settings of Sample 1, refer to the sample
handling and analysis procedures introduced in Chapter 6 Operating Your Analyzer and
prepare the fresh blood samples in the selected count mode to perform the fresh blood
calibration.
Saving calibration results
If non-numeric parameter values ("***") are obtained, a message box will pop up to warn you,
as Figure 9-13 shows.
Figure 9-13 A message box to warn you about the invalid results
9-12