5. SELECT "Lyse test". The analyzer will automatically drain the bath and then dispense
2ml of lyse into the bath;
6. Check the scale to see whether the lyse has reached the expected line (the second from
the bottom). If so, press [ENTER] and the analyzer will automatically flush the bath and
dispense lyse and the test is done;
7. If not, repeat steps 5 and 6 several times. If all the tries have failed, check whether the
lyse has run out or the lyse pickup tube is not properly connected to this analyzer. If the
lyse is still enough and the tube is well connected to the analyzer, contact the Mindray or
your local distributor for assistance.

10.3.9 Clean Bath

Follow the steps given below to perform the "Clean bath" procedure:
SELECT "Clean bath" to start the cleaning procedure, as Figure 10-20 shows. When the
cleaning is done, the screen will return to the initial state;
Maintaining Your Analyzer
Figure 10-19 Bath