Using the QC Programs

8.3 X-B Analysis

One weakness of quality control using control materials is that the control material is usually
assayed only once a shift. If a significant change in calibration occurs, it might go undetected
for the remainder of the shift. The X-B represents the moving average of hematology values
calculated using an algorithm developed by Dr. Brian Bull. The X-B analysis uses the Bull
algorithm to monitor the performance of the analyzer by tracking the average red cell indices
MCV, MCH, and MCHC on the patient samples run through the instrument in batches of 20.
The Target Value for X-B is similar to the assay value for a commercial control. It is derived
from the patient population analyzed on the instrument. The Limit is the acceptable limit of
variation around the target value.
The X-B analysis demands random samples and therefore, the samples categorized by
diseases are not suitable for its use.

8.3.1 QC Editing

Entering the "X-B Table" screen
Press [MENN] to enter the system menu. SELECT "X-B analysis" (Figure 8-19), to enter the
"X-B Table" screen, as Figure 8-20 shows.
Figure 8-19 System menu
8-15