When you have finished editing the desired settings, press [F2] to deactivate the edit boxes.
Running the calibrator
Use the Mindray- specified calibrator. Using calibrator other than the
specified will lead to misleading results.
Refer to the instructions of use of the calibrator for how to store and use the
calibrator.
In the prediluted calibration mode, you cannot dispense diluent from the
analyzer. It is recommended that you prepare at least 7 cups of diluent
before starting calibrating the analyzer in the prediluted mode.
Keep the sample probe tip away from the tube bottom, otherwise the
aspiration volume may be inaccurate.
When the aspiration is done, remove the sample tube only when the sample
probe is out of the tube.
Samples, controls, calibrator and waste are potentially infectious. Wear
proper personal protective equipment (e.g. gloves, lab coat, etc.) and follow
safe laboratory procedures when handling them in the laboratory.
The sample probe tip is sharp and may contain biohazardous materials.
Exercise caution to avoid contact with the probe when working around it.
Do not re-use such disposable products.
In the whole blood mode
1. At "Calibrator" screen, press [F1] to select "Whole Blood" mode;
2. Present a vial of mixed calibrator to the sample probe so that the tip is well into the tube,
and press the aspirate key and the analyzer will start aspirating sample;
3. When you hear the beep and the sample probe is out of the vial, remove the calibrator.
The sample probe will retract into the analyzer and the analysis progress will be
Using the Calibration Programs
9-6