Mindray BC-2800 Operator's Manual page 57

Auto hematology analyzer
Hide thumbs Also See for BC-2800:
Table of Contents

Advertisement

2. Take out the rinse container and place it on or below the countertop;
3. Remove the container cap and insert the tube end that has no connector into the rinse
container and tighten the cap until properly secured, as Figure 4-4 shows;
Figure 4-4 Insert the tube end into the container
4. Locate the blue fitting marked "RINSE" in the lower right corner of the back of the
analyzer.
5. Plug the blue connector of the tube into the fitting and turn it clockwise until properly
secured.
Connecting the lyse container
1. Take out the lyse pickup tube (the one with an orange connector) from the accessory box,
as Figure 4-5 shows;
Installing Your Analyzer
Figure 4-3 Rinse container
4-7