Mindray BC-2800 Operator's Manual page 255

Auto hematology analyzer
Hide thumbs Also See for BC-2800:
Table of Contents

Advertisement

B.5.2 Normal background
Parameter
WBC
RBC
HGB
HCT
PLT
B.5.3 Linearity range
Parameter
9
WBC (10
/L)
12
RBC (10
/L)
HGB (g/L)
9
PLT (10
/L)
B.5.4 Carryover
Parameter
WBC
RBC
HGB
PLT
B.5.5 Reproducibility (with normal-level control)
Parameter
9
WBC (×10
/L)
12
RBC (×10
/L)
HGB (g/L)
MCV (fL)
9
PLT(×10
/L)
B.6
Input/Output Device
Use the specified devices only.
Appendices
Background result
9
≤ 0.3 × 10
/ L
12
≤ 0.03× 10
/ L
≤ 1 g / L
≤ 0.5 %
9
≤ 7 × 10
/ L
Linearity range
0.0 to 99.9
0.0 to 9.99
0 to 300
0 to 999
Carryover
≤ 0.5 %
≤ 0.5 %
≤ 0.5 %
≤ 1 %
Condition
4.0 to 15.0
3.00 to 6.50
100 to 180
70 to 100
200 to 500
B-3
CV(%)
≤ 3.0
≤ 2.0
≤ 2.0
≤ 1.0
≤ 5.0

Advertisement

Table of Contents
loading

Table of Contents