Maintaining Your Analyzer
Figure 10-52 Removing screws
8. Remove the probe from the probe wipe and insert the localizer into the probe wipe from
the bottom, as Figure 10-53 shows;
Figure 10-53 Using localizer
9. Insert the probe into the probe wipe until it reaches the localizer, as Figure 10-54 shows;
Figure 10-54 Inserting sample probe into probe wipe
10. Retighten the screw to fix the probe and remove the localizer to wrap up the adjustment.
10-35