Mindray BC-2800 Operator's Manual page 189

Auto hematology analyzer
Hide thumbs Also See for BC-2800:
Table of Contents

Advertisement

Using the Calibration Programs
If all the parameter values obtained are numeric, a message box will pop up to confirm the
validity of the results, as Figure 9-14 shows. CLICK "Yes" to save the results to the "Fresh
Blood" screen; CLICK "No" to abort the result.
Figure 9-14 A message box to confirm the validity
Repeat the above steps to run the sample 3 to 5 times (5 is recommended) and the analyzer
will automatically calculate the CV and calibration factor, as shows. Note that the CVs should
be within the ranges specified in Table 9-1.
Figure 9-15 Calibration with fresh blood
The calculated calibration factors should be within the 75% to 125%. Any calculated value
that falls out of the calibration range will be flagged with a "*" at the upper right corner. Other
values will not be displayed. In case of an empty calibration factor, try to find out the reason
and if necessary, contact Mindray customer service department or your local distributor.